Part D Technical Data, Measuring Principle and Block Diagram Sect. 10.1

05/2002

IFC 210 E

81

PHDVXULQJDFFXUDF\

Display, digital values, pulse output

F maximum error in % of measured value (not typical values)
v Flow velocity in m/s and ft/s

Reference conditions similar to EN 29 104

Product

water at 10 – 30°C/ 50 – 86°F

Electrical conductivity

> 300 mS/cm

Power supply (rated voltage)

U

N

(± 2%)

Ambient temperature

20 – 22°C

Warm-up time

60 min

Max. calibration equipment error

10

×

smaller than F

Inlet/outlet runs

10

×

DN / 2

×

DN (DN = meter size)

Primary head

properly grounded and centered

Calibrated on EN 17025 accredit to calibration equipment in direct volumetric comparsion.










* IFS 6000 F (DN 2.5 – 4 and 1/10’’ – 1/6’’)
additional error ± 0,3% of MV

Type/Meter size

Maximum error in % of measured value (MV) ...

Curve

DN

mm inch

v

1 m/s /

3 ft/s

v < 1 m/s

DN 2.5 – 6*

1

/

10

’’ –

1

/

4

’’ *

± 0.5% of MV

± (0.4% of MV + 1 mm/s)

B

± (0.4% of MV + 0.04 inch/s

DN 10

3

/

8

’’

± 0.3% of MV

± (0.2% of MV + 1 mm/s)

A

±(0.2% of MV + 0.04 inch/s)

Current output

same error limits as above, additionally ± 10

µ

A

Reproducibility and

repeatability

0,1% of MV, minimum 1 mm/s / 0.04 inch/s at constant flow

External influences

typical values maximum

values

Ambient temperature

Pulse output

0,003% of MV (1)

0,01 % of MV (1)

Current output

0,01 % of MV (1)

0,025% of MV (1)

}

per 1 K / 1.8° F
temperature variation

Power supply

< 0,02 % of MV

0,05 % of MV.

at 10% variation

Load

< 0,01 % of MV

0,02 % of MV

at max. permissible
load, see Sect. 10.1
and 2.5

(1)

All KROHNE signal converters undergo burn-in tests, duration minimum 20 hours at varying ambient temperatures
– 20 to + 60°C/– 4 to + 140°F. The tests are controlled by computers.