beautypg.com

KROHNE IFC 210 E-EEx EN User Manual

Page 49

background image

Part B IFC 210 Signal converter Sect. 5.16

05/2002

IFC 210 E

49

5.16 Primary head – Set data (Fct. 3.02)

Fct. 3.02 Flow Meter
Press

key.

Diameter = set meter size (see instrument nameplate)
Select size from table of meter sizes:
DN 2.5 - 1600 mm
equivalent to

1

/

10

- 64 inch

Select using

or

key.

Transfer to subfunction “Full Scale” with

key.

Full Scale = set full-scale range
Set as described in Sect. 5.1.

Note: if “Pulsvalue” is displayed after pressing

key.

Pulse/Vol. is set under Fct. 1.06 Pulsoutput P, subfunction “Pulsvalue” . Because the full-scale
range Q

100%

has been changed, the output frequency (F) of the pulse output is over- or undershot:

P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Change pulse value accordingly, see Sect. 5.7 Pulseoutput P, Fct. 1.06.

Press

key to change to subfunction “GK Value”.

GK Value = set primary constant GK,

1.0000 - 9.9999

(note information on instrument nameplate, do not change setting !)

Change flashing digit (cursor) with

or

key.

Shift cursor 1 place to right or left with

or

key.

Press

key to change to subfunction “Field Frequency”.

Field Frequency = set magnetic field frequency

1/2

1/18

(1/2, 1/6, 1/18 or 1/36 of power frequency, see instrument nameplate,

1/6

1/36

do not change setting !)

Select with

or

key.

Press

key to change to subfunction “Flow Direction”.

(only for units with DC power supply, transfer to subfunction “Line Frequency”).

Line Frequency = normal line frequency in your country
(Please note, applicable for flowmeters equipped with DC in power supply!)

50 Hz

Select with

or

key.

60 Hz

Press

key to change to subfunction “Flow Direction”.

Flow Direction = set flow direction

+ Direction

(for identification of flow direction, see “+ arrow” on primary head;

– Direction

for F/R mode, identifies the positive flow direction)

Select with - or ¯ key.
Press

key to return to Fct. 3.02 Flow Meter.

Zero check, see Fct. 3.03 in Sect. 5.14 or in Sect. 7.1.

Please refer to Sect. 3.2 Factory settings .



This manual is related to the following products: