KROHNE IFC 210 E-EEx EN User Manual
Page 56
Sect. 6.1 Part C Special applications, functional checks, service, and order numbers
IFC 210 E
05/2002
56
6.1.3 Technical data and terminal assignment
Supply power
Terminals
L or 1L: z28
Plug
XB
N or 0L: z32 Plug
XB
24 DC / AC
U
N
= 24 V DC +30% / -25%, 15 W
U
N
= 24 V AC +10% / -15%, 20 VA
I
N
≤
1.6 AT internal fuse protection
100 ... 230 V AC
U
N
= 100 ... 230 V AC +10% / -15%, 25 VA
I
N
≤
0.8 AT internal fuse protection
Protective conductor/equipotential
bonding
PE / PA
Terminals
d14, d16, d18, d20
Plug XB
d2 ... d12, d22 ... d32
plug XB (optional)
Signalling circuits
Outputs and inputs
Terminals:
Analog
d16, d18, z18
Plug XC:
Binary
d24, d26, z26
Plug XC
Pulses
d32, z32
Plug XC
RS 485
d16, d18, d20, d22, d24
Plug XC (optional)
DC voltage
U
N
(DC)
≤
50 V (connection to devices with
AC voltage
U
N
(AC)
≤
25 V operating voltages up to 250 V)
Field circuit
Terminals
Core 7
d30
Plug XA
Core 8
d32
Plug XA
Frequency-controlled DC voltage
U
N
≤
40 V
Internal fuse protection
I
N
≤
160 mAF
Electrode circuit
in Intrinsic Safety type of protection EEx ib IIC
Terminals:
Signal cores 2, 3
Z6, z10
Plug XA
Shield
20,
30
(BTS
conductor
only)
D6, d10
Plug XA
Shield 1
d8, z8
Plug XA
Outer shield
d2, z2, d4, z4
Plug XA
Peak values
U
0
≤
18 V
(cumulative value)
I
0
≤
40 mA
(cumulative value)
P
0
≤
80 mW
The intrinsically safe electrode circuit is safety-separated (isolated) from the non-intrinsically safe
circuits up to a peak nominal voltage of 375 V.
Ambient temperature
max. allowable –20 to +55°C