KROHNE IFC 210 E-EEx EN User Manual

Page 38

background image

Sect. 5.1

Part B IFC 210 Signal converter

IFC 210 E

05/2002

38

5

'HVFULSWLRQRIIXQFWLRQV

5.1

Full-scale range Q

100%

(Fct. 1.01)

Fct. 1.01 FULL SCALE
Press

key.

Choice of unit for full-scale range Q

100%

m

3

/hr

(cubic metres per hour)

Liter/Sec

(litres per second)

US Gal/min

(US gallons per minute)

User-defined unit, factory-set is „Liter/hr“ (litres per hour) or “US MGal/day”, see Sect. 5.07

Select with

or

key.

Use

key to transfer to numerical setting, 1st number (cursor) flashes..

Set full-scale rangeQ

100%

The setting range is dependent on meter size (DN) and flow velocity (v).

Q

min

= DN

2

×

v

min

Q

max

= DN

2

×

v

max

(refer to flow table in Sect. 10.2)


0.00531 – 86 858 m

3

/hr

0.00147 – 24 120 Liter/Sec

0.02335 – 382 420 US Gal/min


Change flashing number (cursor) with

or

key.

Use

key to shift cursor 1 place to right.

Press

key to return to Fct. 1.01 Full Scale.

Note if “Pulsvalue” is displayed after pressing

key:

Pulse/Vol. is set under Fct. 1.06 Puls B1, subfunction “Select P”. Due to the changed full-scale
range Q

100%,

the output frequency (F) of the pulse output will be over- or undershot:

P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Change pulse value accordingly, see Sect. 5.7 pulse output P, Fct. 1.06.


5.2

Timeconstant (Fct. 1.02)

Fct. 1.02 Timeconstant
Press

key.

Choice

All

(applies to display and all outputs)

Only I

(applies only to display, current and status output)

Select with

or

key.

Transfer to number setting with

key. 1st number (cursor) flashes.

Set numerical value

0.2 - 99.9 Sec (seconds)


Change flashing number (cursor) with

or

key.

Use

key to shift cursor 1 place to right.

Press

key to return to Fct. 1.02 Timeconstant.

π

4

π

4