5.13 Language (Fct. 3.01)

Fct. 3.01 Language
Press

key.

Select language for texts in display

D

(German)

GB/USA

(English)

F

(French)

others on request

Select with

or

key.

Press

key to return to Fct. 3.01 Language.


5.14 Zero check (Fct. 3.03)

Set zero flow in the pipeline, but make sure that the measuring tube is completely filled
with fluid.

Switch on the system and wait 15 minutes.

Press the following keys for zero measurement:

Key Display

Description

If “Yes” set under Fct. 3.04 ENTR Y CODE, key in

9-keystroke CODE 1 now:

Fct. 1.00

Operation

2x

Fct. 3.00

Installation

Fct. 3.01

Language

2x

Fct. 3.03

Zero Point

back

enquire

0.00

- - - - - / - - -

Zero measurement in progress, duration approx. 50 seconds.

When flow > 0 WARNING notice appears, confirm with

key.

Store No

If new value not to be stored, press

key (3x) 4x = return to

measuring

mode.

Store Yes

Fct. 3.03

Zero Point

Store new zero value.

(2x) 3x

- - - - - - -

- - - - - / - - -

Measuring mode with new zero.


5.15 Determine Entry Code (Fct. 3.04)

Fct. 3.04 Entry Code
Press

key.

Choice

No

(no code, enter setting mode with

key)

Yes

(enter setting mode with

key and Code 1:

)

Select with

or

key.

Press

key to return to Fct. 3.04 Entry Code.