KROHNE IFC 210 E-EEx EN User Manual
Page 48

Sect. 5.13
Part B IFC 210 Signal converter
IFC 210 E
05/2002
48
5.13 Language (Fct. 3.01)
Fct. 3.01 Language
Press
→
key.
Select language for texts in display
•
D
(German)
•
GB/USA
(English)
•
F
(French)
•
others on request
Select with
↑
or
↓
key.
Press
↵
key to return to Fct. 3.01 Language.
5.14 Zero check (Fct. 3.03)
•
Set zero flow in the pipeline, but make sure that the measuring tube is completely filled
with fluid.
•
Switch on the system and wait 15 minutes.
•
Press the following keys for zero measurement:
Key Display
Description
→
If “Yes” set under Fct. 3.04 ENTR Y CODE, key in
9-keystroke CODE 1 now:
→
→
→
↵
↵
↵
↑
↑
↑
Fct. 1.00
Operation
2x
↑
Fct. 3.00
Installation
→
Fct. 3.01
Language
2x
↑
Fct. 3.03
Zero Point
→
back
↑
enquire
↵
0.00
- - - - - / - - -
Zero measurement in progress, duration approx. 50 seconds.
When flow > 0 WARNING notice appears, confirm with
↵
key.
Store No
If new value not to be stored, press
↵
key (3x) 4x = return to
measuring
mode.
↑
Store Yes
↵
Fct. 3.03
Zero Point
Store new zero value.
(2x) 3x
↵
- - - - - - -
- - - - - / - - -
Measuring mode with new zero.
5.15 Determine Entry Code (Fct. 3.04)
Fct. 3.04 Entry Code
Press
→
key.
Choice
•
No
(no code, enter setting mode with
→
key)
•
Yes
(enter setting mode with
→
key and Code 1:
→
→
→
↵
↵
↵
↑
↑
↑
)
Select with
↑
or
↓
key.
Press
↵
key to return to Fct. 3.04 Entry Code.