KROHNE IFC 210 E-EEx EN User Manual
Page 51
Part B IFC 210 Signal converter Sect. 5.18
05/2002
IFC 210 E
51
5.18 Application (Fct. 3.06)
Fct. 3.06 Application
Press
→
key.
Flow = Set characterization for the flow
•
steady
flow is steady, max. 150 % of Q
100%
•
pulsating
pulsating flow, = up to 1000 % of Q
100%
, e.g. due to
reciprocating pumps,
Select with
↑
or
↓
key.
see also Sect. 6.4 - 6.6.
Press
↵
key to return to Fct. 3.06 Application.
Empty pipe = signals that measuring tube is empty (possible only with built-in option)
•
No
•
Yes
Select with
↑
or
↓
key.
see also Sect. 6.4 - 6.6.
Press
↵
key to return to Fct. 3.06 Application.
5.19 Hardware (Fct. 3.07)
Assignment of terminals B1 + B2 and field current supply
Fct. 3.07 Hardware
Press
→
key.
Terminal B1 = define function of terminal B1 (valid for Fct. 1.07)
•
Statusoutput
•
Controlinput
Select with
↑
or
↓
key.
Press
↵
key to change to subfunction “Terminal B2”.
Terminal B2 = define function of terminal B2 (valid for Fct. 1.08)
•
Statusoutput
•
Controlinput
Select with
↑
or
↓
key.
Press
↵
key to change to subfunction “Fieldcurrent”.
Field current = define field current supply
•
Intern
•
Extern (> DN 1600 / > 64”)
Select with
↑
or
↓
key.
Press
↵
key to return to Fct. 3.07 Hardware.
Please note:
Only correct settings will give accurate measurement results.
The setting “external” is only required if a field current power driver is connected.