5.11 Set limit values (status output B1 and/or B2)

(Before the following settings can be carried out, set the “Status output” mode for output
terminals B1 and/or B2 in Fct. 3.07, see Sect. 5.19.)

Status output: Fct. 1.07 (output terminal B1) and / or Fct. 1.08 (output terminal B2)

Press

key.

Set Function ”Trip Points“

with

or

key.

Press

key.

Set flow direction:

+ direction

direction

2

directions

Press

key.

Display:

XXX - YYY

Set Trip Point, cursor for the 1st digit shown negatively.

Set with

or

keys..

Use

key to shift cursor 1 place to right.

Setting ranges for XXX and YYY:

0 – 150 % of Q

100%

full-scale range

hysteresis

1%

(=difference between XXX and YYY value)

Press

key to return to Fct. 1.07 or 1.08 Statusoutput.

Switching behaviour (N/O or N/C contact) and hysteresis are adjustable.

N/O contact: XXX- value > YYY- value

N/C contact: XXX- value < YYY- value

Switch closes when flow

exceeds XXX value

Switch opens when flow
exceeds YYY value

Example 1: XXX

=

55%

Example 2: XXX

=

45%

YYY = 45%

YYY = 55%

hysteresis = 10%

hysteresis = 10%

45%

55%

Y

Q

Y Y

X X X

45%

55%

Y

Q

Y

Y

X

X

X

Please note !:

When the two status outputs B1 and B2 are activated (see Sect. 5.19),

min. and max. values can, for example, be signalled.





YYY hysteresis XXX

YYY hysteresis XXX

45% 55% Q

45% 55% Q