5.11 Set limit values (status output B1 and/or B2)
(Before the following settings can be carried out, set the “Status output” mode for output
terminals B1 and/or B2 in Fct. 3.07, see Sect. 5.19.)
Status output: Fct. 1.07 (output terminal B1) and / or Fct. 1.08 (output terminal B2)
•
Press
→
key.
•
Set Function ”Trip Points“
with
↑
or
↓
key.
•
Press
↵
key.
•
Set flow direction:
+ direction
–
direction
2
directions
•
Press
↵
key.
Display:
XXX - YYY
•
Set Trip Point, cursor for the 1st digit shown negatively.
Set with
↑
or
↓
keys..
Use
→
key to shift cursor 1 place to right.
Setting ranges for XXX and YYY:
0 – 150 % of Q
100%
full-scale range
hysteresis
≥
1%
(=difference between XXX and YYY value)
•
Press
↵
key to return to Fct. 1.07 or 1.08 Statusoutput.
•
Switching behaviour (N/O or N/C contact) and hysteresis are adjustable.
N/O contact: XXX- value > YYY- value
N/C contact: XXX- value < YYY- value
Switch closes when flow
exceeds XXX value
Switch opens when flow
exceeds YYY value
Example 1: XXX
=
55%
Example 2: XXX
=
45%
YYY = 45%
YYY = 55%
hysteresis = 10%
hysteresis = 10%
45%
55%
Y
Q
Y Y
X X X
45%
55%
Y
Q
Y
Y
X
X
X
Please note !:
When the two status outputs B1 and B2 are activated (see Sect. 5.19),
min. and max. values can, for example, be signalled.
YYY hysteresis XXX
YYY hysteresis XXX
45% 55% Q
45% 55% Q