beautypg.com

KROHNE IFC 210 E-EEx EN User Manual

Page 35

background image

Part B IFC 210 Signal converter Sect. 4.4

05/2002

IFC 210 E

35

Fct.

Display-Texts

Description and settings

3.06 Application

Set

application

conditions

Flow

Set characterization for flow

• Steady

(steady flow, max. 150% of Q

100%

)

• Pulsating

(pulsating flow, up to 1000% of Q

100%

e.g. caused by reciprocating pumps, refer to Sect. 6.4)

Press

key to transfer to subfunction “Empty Pipe”.

Empty Pipe

Signal when measuring tube is empty (refer to Sect. 6.3 )

• no • yes (possible only with built-in option)

Press

key to return to Fct. 3.06 “Application”.

3.07

Hardware

Assign outputs and inputs to terminals B1 and B2

Terminal B1

Define function of terminal B1 (valid for Fct. 1.07)

• Statusoutput • Controlinput (

Select with

key.

)

Press

key to transfer to subfunction “Terminal B2”.

Terminal B2

Define function of terminal B2 (valid for Fct. 1.08)

• Statusoutput • Controlinput (

Select with

key.

)

Press

key to transfer to subfunction “Fieldcurrent”.

Fieldcurrent

Define field current supply

• Intern

• Extern (> DN 1600 / > 64“ with power driver)

Press

key to return to Fct. 3.07 “Hardware”.

3.08

Location

Set measuring point tag

Factory

setting: ALTOMETER

Characters assignable to each place:

• A-Z, a-z, 0-9 or „-“ (=blank character).)

Press

key to return to Fct. 3.08 “Location”.

3.09

Communication

Set comminucation interface

• Off

(switched off))

HART (HART

®

-interface switched on))

• KROHNE

(KROHNE RS 485-interface switched on),

(only provided if daughter board installed (option).)

• Address:

„HART“ 00-15 / „KROHNE“ 000-239

• Baud rate: 1200, 2400, 4800, 9600 or 19200

(appears with selection “KROHNE” only)“)

Press

key to return to Fct. 3.09 “Communication”.


This manual is related to the following products: