KROHNE IFC 210 E-EEx EN User Manual
Page 19
Part A System installation and start-up Sect. 2.4
05/2002
IFC 210 E
19
2.4
Control inputs B1 and B2
•
The control inputs are galvanically isolated from the current output and all input circuits.
•
Setting data and functions can note down in Section 3.3.
Please also refer to Sect. 3.2 Factory settings.
•
Typical current inputs B1 and B2
B1/B2
B
⊥
•
All operating data and functions can be set,
see Sect. 4 and 5.10, Fct 1.07 and/or 1.08 for operator control
•
The control inputs must be operated in the passive mode.
Function of the control inputs
inactive active
no voltage
voltage present
Off
No functions
External range
High range
Low range
Hold measured values
Measured values follow the Hold measured values
measurement
Measured values at zero
Measured values follow the
measurement
Measured values set to
“zero“
Reset totalizer
inactive Reset
totalizer
Delete errors
inactive
Delete error messages
Connection diagram, see Sect. 2.5: diagram