KROHNE IFC 210 E-EEx EN User Manual
Page 42
Sect. 5.6
Part B IFC 210 Signal converter
IFC 210 E
05/2002
42
5.6
Currentoutput I (Fct. 1.05)
Fct. 1.05 Currentoutput I
Press
→
key.
Function = Select function for current output,
•
Odd
switched off, no function
•
1 Direction
1 flow direction
•
2 Directions
2 flow directions, F/R mode, forward/reverse
Select with
↑
or
↓
key.
Press
↵
key to change to subfunction “Range I”.
Exceptions: When “Off” selected, return to Fct. 1.05 Currentoutput I.
Range I = select measuring range (I
0%
- I
100%
)
•
0 - 20 mA
•
4 - 20 mA
}
fixed ranges
•
mA (user-defined value)
I
0%
– I
100%
(Wert I
0%
< I
100%
!)
0...16
mA
–
4...20
mA
Press
→
key to transfer to number setting.
Select with
↑
or
↓
key.
Press
↵
key to change to subfunction “I Max”.
I Max = Set (I
max
)
•
20.5 mA
•
22 mA
Select with
↑
or
↓
key.
Press
↵
key to change to subfunction “I Error”.
I Error = set error value, (I
Error
)
•
22 mA
(fixed value)
•
0.0 - I
0%
mA
variable value; only variable when I
0%
> 1 mA, see “Range I” above)
Select with
↑
or
↓
key.
Press
↵
key to change to subfunction “Reverse Range”.
Reverse Range = define full-scale range for reverse flow
•
XXX . XX %
5 - 150 % of full-scale range for forward flow
Change negative-image number (cursor) using
↑
or
↓
key,
Move cursor 1 place to right with
→
key.
Press
↵
key to return to Fct. 1.05 Currentoutput I
Please refer to Sect. 3.2 Factory settings .
Refer to Sect. 2.5 for connection diagrams, and to Sect. 5.22 for characteristics.