Error messages in measuring mode

The following list gives all errors that can occur during process flow measurement. Errors shown
in display when “Yes” set in Fct. 1.04 Display, subfunction “Messages.”.

Error messages

Description of error

Error clearance

Line Int.

Power failure Note:
no counting during power failure

Cancel error in Reset-Quit menu, see
Sect. 4.6.
Reset totalizer if necessary.

Overflow I

Current output overranged
(Flow > I Max)

Check and if necessary correct
instrument parameters. After
elimination of cause, error message
deleted automatically.

Overflow P

Pulse output overranged.
(Flow > I Max)

Check and if necessary correct
instrument parameters. After
elimination of cause, error message
deleted automatically.

Totalizer

Totalizer has been reset.

Cancel error message in
Reset/Quit. menu, see Sect. 4.6.

ADC

Analog / digital
converter overranged

Error message deleted automatically
after elimination of cause.

Fatal Error

Fatal error, all outputs set
to “min. values“

Please consult factory.

Empty Pipe

Pipe has run dry.
This message appears only when
the “empty pipe identifier”
option is installed
and the function
is switched on under Fct. 3.06
Application, submenu ”Empty Pipe”.

Fill pipe.