Tx+ tx – KROHNE IFC 210 E-EEx EN User Manual
Page 60
Sect. 6.2 Part C Special applications, functional checks, service, and order numbers
IFC 210 E
05/2002
60
6.2.2 KROHNE RS 485 Interface (Option)
Electrical connection
RS 485 Interface (Option, connection at multipole connector XD)
*
internal connection
Address, e.g.:
001 002 003
*
The shields can be grounded as follows:
1) directly at both ends
or where there is risk of circulating currents,
2) directly at one end (static shielding) or
3) directly at one end and capacitive at the other end.
In the case of 2) and 3), a decrease in EMC is to be expected.
It is essential to blank off the final signal converter's electrical bus. To do this, solder up the
semicircular circuits of solder points S3 and S4 on the RS485 - PCB. For further information refer
to section 8.7.
The RS 485 protocol is available. Please contact your local KROHNE supplier / company.
Settings for use in interface operation
Fct.
Parameter
KROHNE RS 485 Interface
3.09
Communication KROHNE
Address 000-239
Baud
rate
•
1200
•
9600
•
2400
•
19200
•
4800
For further information on setting the signal converter refer to chapters 4 and 5.
TX+
TX-
*
*
*
d16
d18
d20
d22
d24
z16
z18
z20
z22
z24