Part C Special applications, functional checks, service, and order numbers Sect. 7.3

05/2002

IFC 210 E

67

7.3

Hardware information and error status, Fct. 2.02

Before consulting factory about errors or flow measurement problems, please invoke Fct.
2.02 Hardwareinfo (hardware information).

An 8-character and a 10-character status code are stored under this function in
each of 3 (4) “windows”. These 6 (8) status codes allow rapid and simple diagnosis of your
compact flowmeter.

Switch on system.

Press the following keys for display of the status codes:

Key Display

Description

If “Yes” set under Fct. 3.04 ENTR Y CODE, key in

9-keystroke CODE 1 now:

Fct. 1.00

Operation

Fct. 2.00

Test

Fct. 2.01

Test Q

Fct. 2.02

Hardware Information

Modul ADW

-.- - - - -.- - 1st window

- - - - - - - - -

Modul EA

-.- - - - -.- - 2nd window Sample status code

- - - - - - - - -

3.25105.02 ( 8-character code, 1st line)

Modul Anzeige

-.- - - - -.- - 3rd window

3A47F01DB1 (10-character code, 2nd line)

- - - - - - - - -

Modul RS

-.- - - - -.- - 4rd window

(only provided if “computer interface”

- - - - - - - - -

Option installed)

PLEASE NOTE DOWN ALL 6 (8) STATUS CODES !

Fct. 2.02

Hardware Information

Terminate hardware information

(2x) 3x

- - - - - - - -

- - - - - / - - -

Measuring mode

If you need to return your flowmeter to KROHNE, please refer to last

but one page of these Instructions !

7.4

Faults and symptons during start-up and process flow measurement

CAUTION!
The following applies to primary heads in hazardous areas:

Installation, electrical connection, commissioning and maintenance work may only
be carried out by persons who have been trained for work in hazardous areas and
are authorized to do so.

Most faults and symptoms occurring with the flowmeters can be eliminated
with the aid of the following tables.

For greater clarity, faults and symptoms in the tables are divided into various groups.

Groups: D

Display

I

Current output

P

Pulse output

B

Status output or control input

D / I / P / B

Display, current output, pulse output,
binary outputs/inputs

Before contacting KROHNE Service, please read through the

following tables. THANK YOU !


Safety notice
Some of the following measurements are carried out when the system is in the switched on
condition. Exercise caution when handling measuring leads in the vicinity of power connections

electric shock hazard and risk of short-circuiting!