Sect. 7.4 Part C Special applications, functional checks, service, and order numbers

IFC 210 E

05/2002

68

Gruppe D

Display shows . . .

Cause

Remedial action

D 1

Line Int.

Power failure.
Note: no counting during
power failure.

Delete error message in
RESET/QUIT. menu. Reset
totalizer(s) if need be.

D 2

I Overrange

Current output overranged.
(Flow > measuring range)

Check instrument parameters and
correct if necessary. Error message
deleted automatically after cause
has been eliminated.

D 3

P Overrange

Pulse output overranged.
Note: totalizer deviation
possible
(Flow > I Max)

Check instrument parameters,
correct if necessary, and reset
totalizer(s). Error message deleted
automatically after cause has been
eliminated.

D 4

Totalizer

Counts lost
(overflow, data error)

Delete error message in
RESET/QUIT. menu.

D 5

ADC

Analog/digital converter
overranged.

Error message deleted
automatically after cause has been
eliminated.

D 6

Fatal Error

Fatal-Error, all outputs set to
“min.” values.

Replace signal converter
(see Sect. 8.3) or contact KROHNE
Service, having first noted down
hardware information and error
status, see Fct. 2.02.

D 7

Startup
cyclic flashing

Hardware fault,
Watchdog activated.

Replace signal converter
(see Sect. 8.3) or contact
KROHNE Service.

D 8

Pipe empty

The measuring tube is empty.
This message appears only when
the Option ”empty pipe
identifier“
is built in and when the
function is switched on under Fct.
3.06 Application, submenu
“empty pipe“.

Fill the pipe.

D 9

Unsteady display

– Process product

conductivity too low,
particles/air inclusions too large
or inhomogeneous

– Pulsating flow
– Time constant too low or

switched off.

Increase time constant,
see Fct. 1.02, or switch on.

Power supply OFF

Power supply ON

D 10

No display

Check power fuse F5.

Replace if defective
(see Sect. 8.2).

Group I

Fault / Symptom

Cause

Remedial action

Incorrect connection/polarity

Connect properly, see Sect. 2.4.

Receiver instrument or
current output defective.

Check output (see Sect. 7.2) with
new milliammeter: Test ok,
check connection cables and
receiver instrument, replace if
necessary. Test unsuccessfully,
current output defective. Replace
signal converter (see Sect. 8.3) or
contact KROHNE Service

Current output disabled,
see Fct. 1.05

Activate under Fct. 1.05.

Short-circuit between current
output and pulse output.

Check connection and cables,
see Sect. 2.4. Voltage between
I+ and I

(connecting block XC

Pins d16 and z18), approx. 24 V.
Switch off device, eliminate short-
circuit, and switch device on again.

I 1

Receiver instrument
indicates “0”.

External voltage source defective

Check power lines and/or external
voltage source, and replace if
necessary.