Part C Special applications, functional checks, service, and order numbers Sect. 7.4

05/2002

IFC 210 E

69

Group I

Fault / Symptom

Cause

Remedial action

I 2

Unsteady display

– Process product conductivity

too low, particles/air inclusions
too large or inhomogeneous

– Pulsating flow
– Time constant too low or

switched off.

Increase time constant,
see Fct. 1.02

Group P

Fault / Symptom

Cause

Remedial action

Incorrect connection/polarity

Connect properly, see Sect. 2.4.

Totalizer or external voltage
source defective.

Check output (see Sect. 7.2) with
new totalizer: Test ok,
check connection cables and
previous totalizers and external
voltage source, and replace if
necessary. Test faulty,
pulse output defective, replace
signal converter (see Sect. 8.3) or
contact KROHNE Service.

Current output is external
voltage source, short circuit
or current / pulse output defective

Check connection and cables,
see Sect. 2.4. Voltage between
I+ and I

(connecting block XC

Pins d16 and z18), approx. 24 V.
Switch off device, eliminate short-
circuit, and switch device on again.
If no function, then current or pulse
output defective. Replace signal
converter (see Sect. 8.3) or contact
KROHNE Service.

P 1

Totalizer connected but does
not count any pulses

Pulse output is deactivated,
see Fct. 1.06.

Switch on under Fct. 1.06.

P 2

Unsteady pulse rate

– Process product conductivity

too low, particles/air inclusions
too large or inhomogeneous

– Pulsating flow
– Time constant too low or

activated only for current output I

Increase or switch on time
constant, under Fct. 1.02 or switch
on if necessary.

P 3

Pulse rate too high or too
low.

Incorrect setting for pulse output.

Change setting under Fct. 1.06.

Group B

B 1

No function

Faulty connection

Connect properly, see Sect. 2.4.

Output/input defective

Replace electronics unit of signal
converter, see Sect. 8.3

External or internal voltage source
not supplying voltage

Check voltage source, rectify any
short-circuit or replace defective
voltage source.

Receiver instruments faulty

Check connection, if necessary
replace receiver instruments.

Function setting faulty

Correct function settings.

Gruppe D / I / P / B

D / I / P / B 1

Unsteady display and outputs – Process product conductivity

too low, particles/air inclusions
too large or inhomogeneous

– Pulsating flow
– Time constant too low

Increase time constant,
under Fct. 1.02

Power supply OFF

Power supply ON

D / I / P / B 2

No display and
no function of outputs

Check power fuse F5.

Replace if defective, see Sect. 8.2.