Sect. 7.1 Part C Special applications, functional checks, service, and order numbers
IFC 210 E
05/2002
66
7
)XQFWLRQDOFKHFNV
7.1
Zero check
•
Set zero flow in the pipeline, but make sure that the measuring tube is completely filled
with fluid.
•
Switch on the system and wait 15 minutes.
•
Press the following keys for zero measurement:
Key Display
Description
→
If “Yes” set under Fct. 3.04 ENTR Y CODE, key in
9-keystroke CODE 1 now:
→
→
→
↵
↵
↵
↑
↑
↑
Fct. 1.00
Operation
2x
↑
Fct. 3.00
Installation
→
Fct. 3.01
Language
2x
↑
Fct. 3.03
Zero Point
→
back
↑
enquire
↵
0.00
- - - - - / - - -
Zero measurement in progress, duration approx. 50 seconds.
When flow > 0 WARNING notice appears, confirm with
↵
key.
Store No
If new value not to be stored, press
↵
key (3x) 4x = return to
measuring
mode.
↑
Store Yes
↵
Fct. 3.03
Zero Point
Store new zero value.
(2x) 3x
↵
- - - - - - -
- - - - - / - - -
Measuring mode with new zero.
7.2
Test of measuring range Q
•
For this test a measured value can be simulated in the range of -110 to +110 percent of Q
100%
(full-scale range set, see Fct. 1.01 Full Scale).
•
Switch on the system..
•
Press the following keys for this test::
Key Display
Description
→
If “Yes” set under Fct. 3.04 ENTR Y CODE, key in
9-keystroke CODE 1 now:
→
→
→
↵
↵
↵
↑
↑
↑
Fct. 1.00
Operation
↑
Fct. 2.00
Test
→
Fct. 2.01
Test Q
→
sure
no
↑
sure
yes
↵
0
%
Current, pulse and status indication outputs indicate the
corresponding values.
± 10
%
↑
± 50
%
Select using
↑
or
↓
key
± 100
%
± 110
%
↵
Fct. 2.01
Test Q
End of test, actual measured values again present at outputs.
(2x) 3x
↵
- - - - - - -
- - - - - / - - -
Measuring mode