4 table of settable functions – KROHNE IFC 210 E-EEx EN User Manual
Page 30
Sect. 4.4
Part B IFC 210 Signal converter
IFC 210 E
05/2002
30
4.4
Table of settable functions
Abbreviations used
B1/B2
Status output, control input
Q
actual flowrate
DN
Nominal size, meter size
Q
100%
100% flow = full scale range
F
M
Conversion factor volume for any unit,
see Fct. 3.05 “Factor Volume”
Q
max
= DN
2
× v
max
(=max. full-scale range Q
100%
F
max
Highest frequency of pulse output
at v
max
= 12 m/s
/ 40 ft/s
)
F
min
Lowest frequency of pulse output
F
T
Conversion factor time for any unit,
Q
min
= DN
2
× v
min
(=min. full-scale range Q
100%
see Fct. 3.05 “Factor Time”
at v
min
= 0,3 m/s
/ 1 ft/s
)
GK
Primary constant
I
Current output
SMU
Low-flow cutoff for I and P
I
0%
Current at 0% flow
v
Flow velocity
I
100%
Current at 100% flow
v
max
Max. flow velocity (12 m/s / 40 ft/s) at Q
100%
I
max
I Max.
v
min
Min. flow velocity (0.3 m/s / 1 ft/s) at Q
100%
P
Pulse
output
F/R
Forward/Reverse flow at F/R operation
P
max
= F
max
/ Q
100%
P
min
= F
min
/ Q
100%
I
Error
I Error, I
0%
≤
I
Error
≥
I
max
Fct.
Display- Texts
Description and settings
1.00
OPERATION
Operations menu
1.01
FULL SCALE
Full-scale range for flowrate Q
100%
Select unit
•
m
3
/hr
•
Liter/Sec
•
US.Gal/min
•
user unit, factory set is Liter/hr or US MGal/day (see Fct. 3.05)
Press
→
key to transfer to number setting.
Setting
ranges:
The ranges are dependent on the meter size (DN) and the
flow velocity (v):
Q
min
= DN
2
×
v
min
Q
max
= DN
2
×
v
max
Nom. dia. /meter size
v
min
= 0,3 m/s (1 ft/s) v
max
= 12 m/s (40 ft/s)
•
DN 2.5–1600 /
1
/
10
– 64:
0.0053 –
86 859 m
3
/hr
0.0237 –
401 080 US Gal/min
Press
↵
key to return to Fct. 1.01 FULL SCALE.
VALUE P
Pulse value has been changed.
With the old pulse values the output frequency (F)
would have been exceeded or not reached.
P
min
= F
min
/ Q
100%
P
max
= F
max
/ Q
100%
Check new values!
1.02 TIMECONST.
Time
constant
Select::
• ALL (applies to display and all outputs)
• ONLY Current output (only display, current and status outputs)
Press
↵
key to transfer to number setting.
Range:
• 0.2 – 99.9 Sec
Press
↵
key to return to Fct. 1.02 TIMECONST.
1.03
L.F.CUTOFF
Low-flow cutoff (SMU)
• OFF (fixed values: ON = 0.1% / OFF = 0.2%)
PERCENT (variable values)
ON
OFF
1 – 19%
2 – 20%
Press
→
key to transfer to number setting.
Note: Cutoff off value must be greater than cutoff on value.
Press
↵
key to return to Fct. 1.03 L.F. CUTOFF.
π