Part D Technical Data, Measuring Principle and Block Diagram Sect. 10.1
05/2002
IFC 210 E
79
10
7HFKQLFDOGDWD
10.1 IFC 210 E signal converter
5DQJHRIDSSOLFDWLRQ
Flow measurement of liquid products
0RGHRIRSHUDWLRQDQGV\VWHPVWUXFWXUH
Measurement principle
Faraday`s laws of induction
Modularity
Measuring system consisting of
signal converter and separate primary head
Versions – signal converter
IFC 210 E (Standard)
- Standard version with large graphic LC display and
integrated HART
®
interface
IFC 210 E / RS 485 (Option)
- same as standard version, but additionally with
RS 485 interface
IFC 210 E / _ / EEx (Option)
- same as standard version, for operation with
primary heads installed in hazardous areas
Interface module (Option)
- RS 485 / Profibus PA (in preparation)
Versions – primary head
Refer to Technical Data in separate Installation
Instructions
,QSXW
Measured variable
Volumetric flow rate
(electrode voltage from primary head)
Measuring range
Dependent on meter size of primary head
see also Table in Sect. 10.2
Full-scale range
6 Liter/h to 86 860 m
3
/h, corresponding
flow velocity v = 0.3 to 12 m/s
Selectable units
m
3
/h, Liter/s, US Gallons/min or freely selectable
unit, e.g. Liter/day
2XWSXW
Output/input signal
Current-
Pulse-
Status-
Control-
Output output output input
Operation active/
passive
active/
passive
passive Passive
Failure signal (error)
22 mA,
and others
– yes –
Load impedance
max. 800
Ω
– – –
Low
flow
Yes – yes –
F/R mode
Yes
yes
–
–
Other functions
detailed technical data (see below)
Yes
yes
yes
yes