Sect. 10.1 Part D Technical Data, Measuring Principle and Block Diagram

IFC 210 E

05/2002

80


Current output

Function

- all operating data settable

- galvanically isolated from all output and
input circuits

- for active or passive mode

Current:

fixed ranges:

0 - 20 mA and 4 - 20 mA

variable ranges:

for Q = 0% I

0%

= 0 - 16mA

for Q = 100% I

100%

= 4 - 20mA

for Q > 100% I

max

= 22mA

}

adjustable
in 1mA
increments

Active mode

max. 800

load

Error identification

0 / 22 mA and variable

Forward/reverse flow measurment

direction identified via status output

Pulse output

Function

- all operating data settable

- galvanically isolated from current output and

all input circuits

- digital pulse division, interpulse period non-uniform,

- therefore if frequency and cycle meters connected

allow for minimum counting interval::

1000

gate time, totalizer

P

100%

[Hz]

Activ mode

connection: electronic totalizers

Passive mode

connection: electronic or electromechanical totalizers

Electrical data

see connection diagrams in Sect. 2.5

Pulse width

automatic: pulse duty cycle 1:1,
max 10 000 pulses/Sec = 10 kHz

variable: 10 ms - 1 s

1

P

100%

[pulses/s] = f

max

[Hz] =

2x pulse width

Forward/reverse flow measurement

flow direction identified via status output

Status output (passive)

Function

settable as measuring range identification for BA
mode, automatic range change,
Overrange, Low Flow Cutoff, indicator for flow
direction, errors or trip point, change-over contact
(Statusoutput B2 inverse to B1)

Electrical data

see connection diagrams in Sect. 2.5

Control input (passive)

Function

- settable for range change, totalizer reset, error reset,
set outputs to min. values or hold actual output values

- initiate function by low or high control signals

Control signals

U

max

: 24 V AC 32

V

DC (any polarity)

low:

1,4 V

2 V

high:

3 V

4 V

Time constant

0.2 - 99.9 s, adjustable in increments of 0.1 second

Low-flow cutoff

Cutoff on value: 1 - 19 %

Cutoff on value: 2 - 20 %

}

of Q

100%

, adjustable

in 1% increments