beautypg.com

KROHNE IFC 210 E-EEx EN User Manual

Page 62

background image

Sect. 6.4 Part C Special applications, functional checks, service, and order numbers

IFC 210 E

05/2002

62

6.4

Pulsating flow

Important, please note!

Observe the directions and regulations and the electrical data specified in the

EC type test certificate.

In addition to the regulations for heavy-current installations (VDE 0100),
pay particular attention to the requirements specified in EN 60079-14 “Electrical
equipment in hazardous areas”
.

Assembly, installation, commissioning and maintenance work may only be carried
out by ”personnel trained in explosion protection” !



Application
downstream of positive-displacement pumps (reciprocating or diaphragm pumps) without
pulsation dampener

Operator control of the signal converter for the new settings, see Sect. 4 and 5.

To change settings

Fct. 3.2 Field Frequency (change magnetic field frequency)

Stroke frequency less than 80 strokes/min (at max. pump lift): do not change setting.

Stroke frequency 80 – 200 strokes/min (at max. pump lift): change setting to 1/2,

only practical with IFS 5000 F (DN 2.5-100 /

1

/

10

”-4”)

and IFS 4000 F (DN 10, 15, 50-100 /

1

/

10

”,

1

/

2

”, 2”-4”),

Please consult factory where other types and meter sizes are concerned.

Please note: given stroke frequencies close to the tripping point of 80 strokes/min,

additional measuring errors of approx. ± 0.5% of the measured value may occasionally
occur.

Fct. 3.06 Application (adjust overload point of the A/D converter to the application) Change
setting to “pulsating”.

Fct. 1.04 Display (change display presentation of flow)
Change setting to “Trend” to allow better assessment of display unsteadiness.

Fct. 1.02 Timeconstant (change time constant)

Set to “All” and time (t) in seconds.

1000

Recommended:

t [Sec] =

min. strokes / min

Example:

min. number of strokes in operation = 50 strokes/minute

1000

t [Sec] =

50 / min

= 20 Sec

With this setting, the residual ripple of the display will amount to
approx.

±

2% of the measured value. Doubling the time constant will

reduce the residual ripple by a factor of 2.

This manual is related to the following products: