Part B IFC 210 Signal converter
Sect. 5.3
05/2002
IFC 210 E
39
5.3
Low Flow Cutoff (Fct.1.03)
Fct. 1.03 Low Flow Cutoff)
Press
→
key.
Choice
•
Off
fixed tripping point::
ON = 0.1 % / OFF = 0.2 %
•
Percent
variable tripping points:
ON = 1 - 19 % / OFF = 2 - 20 %
Select with
↑
or
↓
key.
Transfer to number setting using
→
key (only if “PERCENT” selected).
1st number (cursor) flashes.
Setting the numerical value when “Percent” selected
•
01 to 19
(cutoff “on” value, left of hyphen)
•
02 to 20
(cutoff “off” value, right of hyphen)
Note: The cutoff “off” value must be greater than the cutoff “on” value.
Change flashing number (cursor) with
↑
or
↓
key.
Use
→
key to shift cursor 1 place to right.
Press
↵
key to return to Fct. 1.03 Low Flow Cutoff.
5.4 Internal
electronic
totalizer
The internal electronic totalizer counts in m
3
, regardless of the unit set under Fct. 1.04,
subfunction “totalizer”.
The counting range is dependent upon the meter size and has been selected such that the
totalizer will count for a minimum of 1 year without overflow:
Meter size
Counting range
DN mm
inch
in m
3
US Gal equivalent
2.5 - 50
1
/
10
- 2
0 - 999 999.99999999
0 - 264 172 052.35800
65 - 200
2
1
/
2
- 8
0 - 9 999 999.9999999
0 - 2 641 720 523.5800
250 - 600
10 - 24
0 - 99 999 999.999999
0 - 26 417 205 235.800
700 -1600
28 - 64
0 - 999 999 999.99999
0 – 264 172 052 358.00
Only part of the totalizer count is shown in the display because it is not possible to output
a 14-digit number. Unit and format of the display are freely selectable, see Fct. 1.04, subfunction
“totalizer” and Sect. 5.5. This determines which part of the count is to be displayed.
Display overflow and totalizer overflow are independent of one another.
Example
Internal count
0000123 . 7654321 m
3
Format, display unit
XXXX . XXX
Liter
Internal count in unit
0123765 . 4321000 Liter
Displayed
3765 . 432
Liter