Technical data – KROHNE IFC 100 Converter EN User Manual
Page 80
8
TECHNICAL DATA
80
IFC 100
www.krohne.com
07/2010 - 4000041004 - MA IFC 100 R04 en
Electrical connection
General
Electrical connection is carried out in conformity with the VDE 0100 directive
"Regulations for electrical power installations with line voltages up to 1000 V" or
equivalent national regulations.
Power supply
100…230 VAC (-15% / +10%), 50/60 Hz; non-Ex: standard; Ex: optional
240 VAC + 5% is included in the tolerance range.
12...24 VDC (-55% / +30%); only available in non Ex version
12 VDC - 10% is included in the tolerance range.
24 VAC/DC (AC: -15% / +10%; DC: -25% / +30%); only available as Ex version
12 V is not
not
not
not included in the tolerance range.
Power consumption
AC: 7 VA
DC: 4 W
Signal cable
Only necessary for remote device versions.
DS 300 (type A)
DS 300 (type A)
DS 300 (type A)
DS 300 (type A)
Max. length: 600 m / 1950 ft (depending on electrical conductivity and version of
measuring sensor)
Cable entries
Standard: M20 x 1.5 (8...12 mm)
Option: ½" NPT, PF ½
Outputs
General
All outputs are electrically isolated from each other and from all other circuits.
All operating data and output values can be adjusted.
Description of used abbreviations U
ext
= external voltage; R
L
= load + resistance;
U
o
= terminal voltage; I
nom
= nominal current
Current output
Current output
Current output
Current output
Output data
Volume flow, mass flow, diagnostics value, flow velocity, coil temperature,
conductivity
Settings
Without HART
Without HART
Without HART
Without HART
®
Q = 0%: 0…20 mA; Q = 100%: 10…21.5 mA
Error identification: 0…22 mA
With HART
With HART
With HART
With HART
®
Q = 0%: 4…20 mA; Q = 100%: 10…21.5 mA
Error identification: 3.5…22 mA
Operating data
Operating data
Operating data
Operating data
Active
U
int, nom
= 20 VDC
I ≤ 22 mA
R
L
≤ 750 Ω
Passive
U
ext
≤ 32 VDC
I ≤ 22 mA
U
0
≤ 2 V at I = 22 mA
R
L
≤ (U
ext
- U
0
) / I
max
.book Page 80 Friday, July 9, 2010 12:29 PM