6

OPERATION

68

IFC 100

www.krohne.com

07/2010 - 4000041004 - MA IFC 100 R04 en

6.3.4 Set free units

C5.8 HART

C5.8

HART

This function is only available for devices with a HART

®

interface!

C5.8.1

HART

Switch HART

®

communication on/off

Select:
on (HART

®

active) current = 4…20 mA /

off (HART

®

not active) current = 0…20 mA

C5.8.2

address

Set address for HART

®

operation.

Select: 00 (point-to-point operation, current output has normal function,

current = 4…20 mA) / 01...15 (Multi-Drop operation, current output has a

constant setting of 4 mA)

C5.8.3

message

Set required text:

A…Z ; a…z ; 0…9 ; / - + , . *

C5.8.4

description

Set required text:

A…Z ; a…z ; 0…9 ; / - + , . *

C5.9 quick setup

C5.9

quick setup

Activate quick access in quick setup menu; default setting: quick setup is

active (yes)

Select: yes (activated) / no (not activated)

C5.9.1

reset counter 1

Reset counter 1 in quick setup menu

Select: yes (activated) / no (not activated)

C5.9.2

reset counter 2

Reset counter 2 in quick setup menu

Select: yes (activated) / no (not activated)

C5.9.4

process input

Activate quick access to the important process input parameters

Select: yes (activated) / no (not activated)

Free units

Sequences to set texts and factors

Texts

Texts

Texts

Texts

Volume flow, mass flow and density:

3 digits before and after the slash

xxx/xxx (max. 3 digits before / after the slash)

Volume, mass:

xxx (max. 3 digits)

Permissible characters:

A…Z ; a…z ; 0…9 ; / - + , .*; @ $ % ~ () [] _

Conversion factors

Conversion factors

Conversion factors

Conversion factors

Desired unit

= [unit see above] * conversion factor

Conversion factor

Max. 9 digits

Shift decimal point:

↑ to the left and ↓ to the right