beautypg.com

Operation – KROHNE IFC 100 Converter EN User Manual

Page 60

background image

6

OPERATION

60

IFC 100

www.krohne.com

07/2010 - 4000041004 - MA IFC 100 R04 en

C2.….6

polarity

Set polarity, please note flow direction in C1.2.2!

Select: both polarities (plus and minus values are displayed) / positive

polarity (display for negative values = 0) / negative polarity (display for

positive values = 0) / absolute value (use for the output)

C2.….7

limitation

Limitation before applying the time constant.

±

xxx … ±xxx%; range: -150…+150%

C2.….8

low flow cutoff

Sets output value to "0":

x.xxx ± x.xxx%; range: 0.0…20%

(1st value = switching point / 2nd value = hysteresis),

condition: 2nd value ≤ 1st value

C2.….9

time constant

Range: 000.1…100 s

C2.….10

invert signal

Select:

off (activated output generates a high current at the output, switch closed) /

on (activated output generates a low current at the output, switch open)

C2.….12

information

Serial number of the I/O circuit board, software version number and

production date of the circuit board

C2.….13

simulation

Sequence see B1.… frequency out X

C2.… pulse output X

C2.…

pulse output X

X stands for connection terminals D

…

stands for Fct. no. C2.5 (D)

C2.….1

pulse shape

Specify the pulse shape.

Select: symmetric (about 50% on and 50% off) / automatic (constant pulse

with about 50% on and 50% off at 100% pulse rate) / fixed (fixed pulse rate,

setting see below Fct. C2.….3 100% pulse rate)

C2.….2

pulse width

Only available if set to "fixed" in Fct. C2.….1

Range: 0.05…2000 ms

Note: max. setting value Tp [ms] ≤ 500 / max. pulse rate [1/s], gives the pulse

width = time where the output is activated

C2.….3

max. pulse rate

Pulse rate for 100% of the measuring range.

Range: 0.0…10000 1/s

Limitation 100% pulse rate ≤ 100/s: I

max

≤ 100 mA

Limitation 100% pulse rate > 100/s: I

max

≤ 20 mA

C2.….4

measurement

Measurements for activating the output.

Select: volume flow / mass flow

C2.….5

pulse value unit

Selection of the unit from a list, depending on the measurement.

C2.….6

value p. pulse

Set value for volume or mass per pulse.

xxx.xxx, range in [l] or [kg] (volume or mass for current output C2.….6)

At max. pulse rate see above 2.….3 pulse output.

C2.….7

polarity

Set polarity, please note flow direction in C1.2.2!

Select: both polarities (plus and minus values are displayed) / positive

polarity (display for negative values = 0) / negative polarity (display for

positive values = 0) / absolute value (use for the output)

C2.….8

low flow cutoff

Sets output value to "0"

x.xxx ± x.xxx%; range: 0.0…20%

(1st value = switching point / 2nd value = hysteresis),

condition: 2nd value ≤ 1st value

C2.….9

time constant

Range: 000.1…100 s

No.

Function

Settings / descriptions

.book Page 60 Friday, July 9, 2010 12:29 PM