Operation – KROHNE IFC 100 Converter EN User Manual
Page 73
OPERATION
6
73
IFC 100
www.krohne.com
07/2010 - 4000041004 - MA IFC 100 R04 en
Simulation of the measured values
Information
Messages on the display Description
Actions
Status: C _ _ _ _ _
Output values partially simulated or
fixed
Maintenance required.
C checks in progress
Test mode of the device. Measured values
are possibly simulated values or values with
fixed settings.
Message depending on the situation via
HART
®
or FDT.
C test sensor
Test function of the measuring sensor
electronics is active.
-
C simulation fieldbus
Values on the Foundation Fieldbus interface
are simulated.
-
Messages on the display Description
Actions
Status: I _ _ _ _ _
Information (current measurement OK)
I counter 1 stopped
This is counter 1 or FB2 (with Profibus). The
counter has stopped.
If counter to continue counting, activate
"yes" in Fct. C2.y.9 (start counter).
I counter 2 stopped
This is counter 2 or FB3 (with Profibus). The
counter has stopped.
I power fail
The device was not in operation for an
unknown period of time, because the power
was switched off. This message is for
information only.
Temporary power failure. Counters did not
run during it.
I over range display 1
1st line on page 1 (2) of display limited by
filter setting.
Menu display Fct. C4.3 and/or C4.4, select
1st or 2nd meas. page and increase values
in functions C4.z.3 range and/or C4.z.4
limitation.
I backplane sensor
The data on the backplane are not usable
because they have been generated with an
incompatible version.
-
I backplane settings
The global settings on the backplane are not
usable because they have been generated
with an incompatible version.
-
I backplane difference
The data on the backplane differ from the
data in the display. If the data are usable, a
dialogue is indicated in the display.
-
I write cycles overfl.
The maximum number of write cycles of the
EEPROM or FRAMS on the Profibus DP PCB
has been exceeded.
-
I baudrate search
The baudrate of the Profibus DP interface is
searched for.
-
I no data exchange
There is no data exchange between the
signal converter and the Profibus.
-
I conductivity off
Conductivity measurement switched off.
Changing of settings in Fct. C1.3.1.
I diagnosis channel off
Diagnosis value switched off.
Changing of settings in Fct. C1.3.17.