ELECTRICAL CONNECTIONS

22

IFC 100

www.krohne.com

07/2010 - 4000041004 - MA IFC 100 R04 en

4.4.3 Length of signal cable A

INFORMATION!
For temperatures of the medium above 150°C / 300°F, a special signal cable and a ZD
intermediate socket are necessary. These are available including the changed electrical
connection diagrams.

Measuring sensor

Nominal size

Min. electrical
conductivity
[µS/cm]

Curve for signal
cable A

DN [mm]

[inches]

OPTIFLUX 1000 F

10...150

3/8...6

5

A1

OPTIFLUX 2000 F

25...150

1...6

20

A1

200...1200

8...48

20

A2

OPTIFLUX 4000 F

2.5...150

1/10...6

1

A1

200...1200

8...48

1

A2

OPTIFLUX 5000 F

2.5...100

1/10...4

1

A1

150...250

6...10

1

A2

OPTIFLUX 6000 F

2.5...150

1/10...6

1

A1

WATERFLUX 3000 F

25...600

1...24

20

A1

Figure 4-3: Maximum length of signal cable A

1 Maximum length of signal cable A between the measuring sensor and signal converter [m]
2 Maximum length of signal cable A between the measuring sensor and signal converter [ft]
3 Electrical conductivity of the medium being measured [μS/cm]