ELECTRICAL CONNECTIONS

4

29

IFC 100

www.krohne.com

07/2010 - 4000041004 - MA IFC 100 R04 en

Connecting the signal and field current cables

Connect the electrical conductors as follows:

1 Push the lever downwards with a screwdriver in good condition (blade: 3.5 mm wide and

0.5 mm thick).

2 Insert the electrical conductor into the plug.
3 The conductor will be clamped as soon as the lever is released.

Figure 4-8: Function of the electrical connection terminal

Figure 4-9: Connecting the signal and field current cables

1 Cable entry for field current cable
2 Cable entry for signal cable
3 Connecting the field current cable shield
4 Electrical conductor (7)
5 Electrical conductor (8)
6 Stranded drain wire (1) of the inner shield (10) of the signal cable
7 Electrical conductor (2)
8 Electrical conductor (3)
9 Stranded drain wire (S) of the outer shield (60)

.book Page 29 Friday, July 9, 2010 12:29 PM