OPERATION

6

53

IFC 100

www.krohne.com

07/2010 - 4000041004 - MA IFC 100 R04 en

6.3.2 Menu B, test

No.

Function

Settings / descriptions

B1 simulation

B1

simulation

Displayed values are simulated.

B1.1

flow speed

Simulation of the flow speed

Select: break (exit function without simulation) /

set value (range: -12…+12 m/s; unit selection in Fct. C5.7.7)

Query: start simulation?

Settings: no (exit function without simulation) / yes (start simulation)

B1.2

volume flow

Volume flow simulation, sequence and settings similar to B1.1, see above!

X stands for one of the connection terminals A, C or D

…

stands for Fct. no. B1.3...1.6

B1.…

current output X

simulation X

X stands for one of the connection terminals A, C or D

Sequence and settings similar to B1.1, see above!

For pulse output a set number of pulses are output in 1 s!

B1.…

pulse output X

B1.…

frequency out X

B1.…

limit switch X

B1.…

status output X

B2 actual values

B2

actual values

Display the current values; exit the displayed function with key ^.

B2.1

operating hours

Display the actual operating hours; exit the displayed function with key ^.

B2.2

act. flow speed

Display the actual flow speed; exit the displayed function with key ^.

B2.3

act. coil temp.

See also Fct. C1.1.7...C1.1.8

B2.4

electr. temperature

Display the actual electronics temperature; exit the displayed function with

key ^.

B2.5

act. conductivity

See also Fct. C1.3.1...C1.3.2

B2.6

act. electr. noise

See also Fct. C1.3.13...C1.3. 15

B2.8

act. coil resistance

Display the actual resistance of the field coils depending on the current coil

temperature.

.book Page 53 Friday, July 9, 2010 12:29 PM