Operation – KROHNE IFC 100 Converter EN User Manual
Page 53
OPERATION
6
53
IFC 100
www.krohne.com
07/2010 - 4000041004 - MA IFC 100 R04 en
6.3.2 Menu B, test
No.
Function
Settings / descriptions
B1 simulation
B1
simulation
Displayed values are simulated.
B1.1
flow speed
Simulation of the flow speed
Select: break (exit function without simulation) /
set value (range: -12…+12 m/s; unit selection in Fct. C5.7.7)
Query: start simulation?
Settings: no (exit function without simulation) / yes (start simulation)
B1.2
volume flow
Volume flow simulation, sequence and settings similar to B1.1, see above!
X stands for one of the connection terminals A, C or D
stands for Fct. no. B1.3...1.6
B1.
current output X
simulation X
X stands for one of the connection terminals A, C or D
Sequence and settings similar to B1.1, see above!
For pulse output a set number of pulses are output in 1 s!
B1.
pulse output X
B1.
frequency out X
B1.
limit switch X
B1.
status output X
B2 actual values
B2
actual values
Display the current values; exit the displayed function with key ^.
B2.1
operating hours
Display the actual operating hours; exit the displayed function with key ^.
B2.2
act. flow speed
Display the actual flow speed; exit the displayed function with key ^.
B2.3
act. coil temp.
See also Fct. C1.1.7...C1.1.8
B2.4
electr. temperature
Display the actual electronics temperature; exit the displayed function with
key ^.
B2.5
act. conductivity
See also Fct. C1.3.1...C1.3.2
B2.6
act. electr. noise
See also Fct. C1.3.13...C1.3. 15
B2.8
act. coil resistance
Display the actual resistance of the field coils depending on the current coil
temperature.
.book Page 53 Friday, July 9, 2010 12:29 PM