beautypg.com

Operation – KROHNE IFC 100 Converter EN User Manual

Page 52

background image

6

OPERATION

52

IFC 100

www.krohne.com

07/2010 - 4000041004 - MA IFC 100 R04 en

A5 digital outputs

A5

digital outputs

Valid for pulse output (terminal D) and counter 1.

A5.1

measurement

1) Select measurement: volume flow / mass flow

2) Use for all outputs? (also use this setting for Fct. A5.2...A5.4!)

Setting: no (only for pulse output D) / yes (for all digital outputs)

A5.2

pulse value unit

Selection of the unit from a list, depending on the measurement.

A5.3

value p. pulse

1) Setting for pulse output D (volume or mass value per pulse)

Setting: xxx.xxx in l/s or kg/s

2) Use for all outputs? Make setting, see Fct. A5.1 above!

A5.4

low flow cutoff

1) Setting for pulse output D (sets output value to "0")

Setting: x.xxx ± x.xxx% (range: 0.0…20%)

(1st value = switching point / 2nd value = hysteresis),

condition: 2nd value ≤ 1st value

2) Use for all outputs? Make setting, see Fct. A5.1 above!

process input

A7.1

device serial no.

Serial no. of the system.

The following process input parameters are only available, if the quick access has been activated in the menu "setup /

device / quick setup.

A7.2

zero calibration

Display of current zero calibration value.

Query: calibrate zero?

Settings see Fct. C1.1.1.

A7.3

size

Select from size table.

A.7.5

GKL

Set value acc. to nameplate; range: 0.5…12 (20)

A7.6

coil resistance Rsp

Field coil resistance at 20°C; range: 10.00...220 Ω

A7.7

calib. coil temp.

The coil temperature is derived from the coil resistance at the reference

temperature.

Settings see Fct. C1.1.8.

A7.8

target conduct.

Reference value for on-site calibration; range: 1.000...50000 μS/cm

A7.9

EF electr. factor

For calculation of the conductivity based on the electrode impedance.

Settings see Fct. C1.1.11.

A7.10

field frequency

Setting as on measuring sensor nameplate =

line frequency x value (from the following list):

2; 4/3; 2/3; 1/2; 1/4; 1/6; 1/8; 1/12; 1/18; 1/36; 1/50

A7.11

flow direction

Define polarity of flow direction.

forwards (according to the arrow on the measuring sensor) or backwards (in

the opposite direction to the arrow)

.book Page 52 Friday, July 9, 2010 12:29 PM