Measurements out of specification

Messages on the display Description

Actions

Status: S _ _ _ _ _

Out of specification, measurement
continues, accuracy possibly less.

Maintenance required.

S uncertain measurement Device maintenance necessary; measured

values only conditionally usable.

Group message, when errors as described

below or other influences occur.

S empty pipe

1 or 2 measuring electrodes are not in

contact with the medium; measured value is

not set to zero. Measurement continues.

Filling level of EMF less than 50% or

electrodes completely insulated. If "0" to be

indicated when pipe is empty, activate in

Fct. C1.3.1 "cond.+empty pipe [F]".

The two empty pipe messages cannot appear at the same time. The difference lies in

whether the measured value is also set to zero upon the detection of an empty pipe. The

sensor electronics will use one or the other function (setting to zero or further

measurement) depending on a selection made by the user.

S electrode noise

Noise on the electrodes too high. Measured

values are still supplied. No message if

empty pipe.

a) Electrodes extremely soiled;

b) Conductivity too low: activate noise or

pulse filter Fct. C1.2.4, C1.2.7;

c) Gas bubbles, solids or chem. reactions in

medium: activate noise or pulse filter Fct.

C1.2.4, C1.2.7;

d) Electrode corrosion (if message also

appears when flow is zero): use sensor with

suitable electrode material.

S gain error

Preamplifier not equal to the calibrated

value; check calibration. Measured values

are still supplied.

Defective, replace electronic unit.

S field coil broken

Field coil resistance too high.

Check field coil connections to the

electronic module (for remote versions:

field current cable) for open circuit / short

circuit

S field coil bridged

Field coil resistance too low.

S field current deviation

Measured field current not equal to the

calibrated value. Check calibration.

Measured values are still supplied. No

message if coil broken or bridged.

Check field current connections. If OK:

defective, replace electronic unit.

S electronic temperature

Upper limit for the permissible electronic

temperature has been exceeded.

Ambient temperature too high, direct solar

radiation or, for C version, process

temperature too high.

S coil temperature

Upper limit for the permissible coil

temperature has been exceeded. No

message if coil broken/bridged.

Process and ambient temperature too high.

S overflow counter 1

This is counter 1 or FB2 (with Profibus).

Counter has overrun and started again at

zero.

-

S overflow counter 2

This is counter 2 or FB3 (with Profibus).

Counter has overrun and started again at

zero.

-

S backplane invalid

The data record on the backplane is invalid.

The CRC check has revealed a fault.

No data can be loaded from the backplane

when replacing electronics. Save the data to

the backplane again (Service).