ELECTRICAL CONNECTIONS

44

IFC 100

www.krohne.com

07/2010 - 4000041004 - MA IFC 100 R04 en

4.11.4 HART

®

connection

HART

®

connection active (point-to-point)

The parallel resistance to the HART

®

communicator must be R ≥ 230 Ω.

HART

®

connection passive (Multi-Drop operation)

• I: I

0%

≥ 4 mA

• Multi-Drop mode I: I

fix

≥ 4 mA = I

0%

• U

ext

≤ 32 VDC

• R ≥ 230 Ω

INFORMATION!

The current output at connection terminals A+/A-/A always has HART

®

capability.

Figure 4-20: HART

®

connection active (I

a

)

1 HART

®

communicator

Figure 4-21: HART

®

connection passive (I

p

)

1 HART

®

communicator

2 Other HART

®

- capable devices