OPERATION

6

55

IFC 100

www.krohne.com

07/2010 - 4000041004 - MA IFC 100 R04 en

6.3.3 Menu C, setup

No.

Function

Settings / descriptions

C1 process input

C1.1 Calibration

C1.1

calibration

Grouping of all functions related to measuring sensor calibration.

C1.1.1

zero calibration

Display of actual zero calibration value.

calibrate zero?

Select: break (return with ^ key) /

standard (with factory setting) /

manual (display last value, set new value, range: -1.00…+1 m/s) /

range: -1.00…+1 m/s) /

automatic (shows the current value as the new zero calibration value)

C1.1.2

size

Select from size table.

C1.1.5

GKL

Set value acc. to nameplate; range: 0.5…12 (20)

C1.1.7

coil resistance Rsp

Field coil resistance at 20°C; range: 10.00...220 Ω

C1.1.8

calib. coil temp.

The coil temperature is derived from the coil resistance at the reference

temperature.

Set coil temperature:

break (return with ^ key)

standard (= 20°C)

automatic (set current temperature); range: -40.0…+200°C

Set coil resistance:

break (return with ^ key)

standard (= setting from Fct. C1.1.7)

automatic (= calibration with the current resistance)

C1.1.9

density

Calculation of mass flow with constant density of product;

range: 0.1…5 kg/l

C1.1.10

target conduct.

Reference value for on-site calibration; range: 1.000...50000 μS/cm

C1.1.11

EF electr. factor

For calculation of the conductivity based on the electrode impedance.

Query: calibrate EF?

Select: break (return with ^ key) /

standard (with factory setting) /

manual (set desired value) /

automatic (determines EF according to the setting in Fct. C1.1.10)

C1.1.13

field frequency

Setting as on measuring sensor nameplate =

line frequency x value (from the following list):

1/2; 1/4; 1/6; 1/8; 1/12; 1/18; 1/36; 1/50

C1.1.14

select settling

Select settling (special function)

Select: standard (fixed allocation) /

manual (manual time setting for the settling time for the field current)

C1.1.15

settling time

Only when "manual" selected under Fct. C1.1.14; range: 1.0…250 ms

C1.1.16

line frequency

Set line frequency.

automatic (measuring & setting; for DC systems fixed setting 50 Hz)

Select: 50 Hz or 60 Hz (fixed setting)

C1.1.17

act. coil resistance

Display of the actual resistance of the field coil for calculation of the

temperature.