Operation – KROHNE IFC 100 Converter EN User Manual
Page 55
OPERATION
6
55
IFC 100
www.krohne.com
07/2010 - 4000041004 - MA IFC 100 R04 en
6.3.3 Menu C, setup
No.
Function
Settings / descriptions
C1 process input
C1.1 Calibration
C1.1
calibration
Grouping of all functions related to measuring sensor calibration.
C1.1.1
zero calibration
Display of actual zero calibration value.
calibrate zero?
Select: break (return with ^ key) /
standard (with factory setting) /
manual (display last value, set new value, range: -1.00…+1 m/s) /
range: -1.00…+1 m/s) /
automatic (shows the current value as the new zero calibration value)
C1.1.2
size
Select from size table.
C1.1.5
GKL
Set value acc. to nameplate; range: 0.5…12 (20)
C1.1.7
coil resistance Rsp
Field coil resistance at 20°C; range: 10.00...220 Ω
C1.1.8
calib. coil temp.
The coil temperature is derived from the coil resistance at the reference
temperature.
Set coil temperature:
break (return with ^ key)
standard (= 20°C)
automatic (set current temperature); range: -40.0…+200°C
Set coil resistance:
break (return with ^ key)
standard (= setting from Fct. C1.1.7)
automatic (= calibration with the current resistance)
C1.1.9
density
Calculation of mass flow with constant density of product;
range: 0.1…5 kg/l
C1.1.10
target conduct.
Reference value for on-site calibration; range: 1.000...50000 μS/cm
C1.1.11
EF electr. factor
For calculation of the conductivity based on the electrode impedance.
Query: calibrate EF?
Select: break (return with ^ key) /
standard (with factory setting) /
manual (set desired value) /
automatic (determines EF according to the setting in Fct. C1.1.10)
C1.1.13
field frequency
Setting as on measuring sensor nameplate =
line frequency x value (from the following list):
1/2; 1/4; 1/6; 1/8; 1/12; 1/18; 1/36; 1/50
C1.1.14
select settling
Select settling (special function)
Select: standard (fixed allocation) /
manual (manual time setting for the settling time for the field current)
C1.1.15
settling time
Only when "manual" selected under Fct. C1.1.14; range: 1.0…250 ms
C1.1.16
line frequency
Set line frequency.
automatic (measuring & setting; for DC systems fixed setting 50 Hz)
Select: 50 Hz or 60 Hz (fixed setting)
C1.1.17
act. coil resistance
Display of the actual resistance of the field coil for calculation of the
temperature.