Electrical connections – KROHNE IFC 100 Converter EN User Manual
Page 21
ELECTRICAL CONNECTIONS
4
21
IFC 100
www.krohne.com
07/2010 - 4000041004 - MA IFC 100 R04 en
4.4.2 Preparing signal cable A, connection to signal converter
• Connection of the two shields in the signal converter is carried out via the stranded drain
wires.
• Bending radius: ≥ 50 mm / 2"
Required materials
• PVC insulating tube, Ø2.5 mm / 0.1"
• Heat-shrinkable tubing
• 2x wire end ferrules to DIN 46 228: E 1.5-8 for the stranded drain wires (1, 6)
• 2x wire end ferrules to DIN 46 228: E 0.5-8 for the insulated conductors (2, 3)
1 Strip the conductor to dimension a.
2 Cut off the inner shield (10) and the outer shield (60). Make sure not to damage the stranded
drain wires (1, 6).
3 Slide the insulation tubing over the stranded drain wires (1, 6).
4 Crimp the wire end ferrules onto the stranded drain wire.
5 Crimp the wire end ferrules onto the conductors (2, 3).
6 Pull the heat-shrinkable tubing over the prepared signal cable.
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
Figure 4-2: Preparation of signal cable A
a = 80 mm / 3.15"
.book Page 21 Friday, July 9, 2010 12:29 PM