FIGURE B-5. USING LIST COMMANDS AND QUERIES (SHEET 2 OF 2)

B.53 [SOURce:]LIST:COUNt:SKIP? QUERY

LIST:COUN:SKIP?

Syntax:

Short Form: LIST:COUN:SKIP?

Long Form: LIST:COUNt:SKIP?

Return Value:

Description: Identifies how many steps will skipped the first time the list is executed. Returns value set by

LIST:COUN:SKIP command. (See example, Figure B-5.)

B.54 [SOURce:]LIST:CURRent COMMAND

LIST:CURR

Syntax:

Short Form: LIST:CURR , , . . . (to max of 5900 data points for global dwell
time)
Long Form: LIST:CURRent , , . . . (to max of 5900 data points for global
dwell time)
= digits with decimal point and Exponent, e.g., 2.71E1 for 27.1

Description: Adds the current value (in Amps) to list. This command sequentially adds LIST:CURRent values to

the main channel List Data Table locations illustrated in Table B-2. Starting location is indicated by
LIST:CURR:POIN? The maximum number of entries is 2950,3933 or 5900. Since the input buffer of
the BOP has a limit of 253 characters, multiple commands are necessary to complete the full 3933
entries of the list. If LIST:VOLT has any entries, an error message: -221,”Settings conflict” is posted in
the error queue. Related Commands: LIST:CURR:POIN?. (See example, Figure B-5.).

.VOLT:MODE LIST

Executes the list. For 240 milliseconds the BOP outputs a staircase triangle
wave from -20V to +20V and back down to -20V. This staircase will have a
uniform spacing between voltage changes of 10 milliseconds and will
repeat 100 times.

VOLT?

Returns +20 (the last step in the list set the unit to +20V.

LIST:COUNT?

Returns 100.

LIST:COUNT 10

Determines that the list will be repeated 10 times when executed.

LIST:COUNT?

Returns 10.

VOLT:MODE LIST

Initiates execution of the list. The unit outputs a sawtooth waveform that
increases from -20V to +20V and repeats 10 times. There is an abrupt
change from +20V to -20V each time the list transitions from location 20 to
location 0.

VOLT?

Returns 20, (the last step in the list set the unit to 20V.

LIST:DWELL.01,.01,.01,.01,.01,.01,.01,.01,.01

Enters dwell time of 0.01 seconds in locations 1 through 9.

LIST:DWELL .03

Enters dwell time of 0.03 second in location 10.

VOLT:MODE LIST

the list is not executed, the unit returns error -221,Settings Error indicating
the lists are not balanced.

LIST:DWELL:POINTS?

Returns 11 to indicate the dwell list has 11 entries

LIST:VOLT:POINTS?

Returns 21 to indicate the voltage list has 21 entries.

LIST:DWELL .01,.01,.01,.01,.01,.01,.01,.01,.01,.01

Dwell times are entered in locations 11 through 20.

DWELL:POINTS?

Returns 21 indicating there are now 21 step dwell times on the list.

LIST:QUER?

Returns 0 indicating list queries will start from location 0

LIST:QUER 9

Causes list queries to start at location 9

LIST:DWEL?

Returns 01,.03,.01,.01,.01,.01,.01,.01,.01,.01,.01,.01 (the dwell times for
locations 9 through 20.

LIST COUNT 0

List will repeat indefinitely.

VOLT:MODE LIST

The list is executed. Stair case wave now consists of 21 steps of .01 sec-
onds except for the 0V step (location 10) which outputs 0V for 0.03 second.

LIST:DWEL?

Returns error -221,Settings Error because a LIST command cannot be exe-
cuted while the LIST is still being executed.

VOLT:MODE FIX

Stops execution of the list

LIST:VOLT:APPLY SINE,60,20 Adds a 60Hz, 20V p-p sine wave to the current waveform
VOLT:MODE:LIST

Unit outputs the staircase and then the sine wave.

*RST

Unit stops generating the waveform.