The Master calibration should never be overwritten. Factory, Master, and First are set to the
same values when the BOP is calibrated at the factory. The Working calibration is the active cal-
ibration. Each time a CAL:SAV is executed, the values are saved in the Working (active) area.
At the same time, the values previously stored in Working are moved to Prior, and the values
previously stored in Prior are moved to Oldest. The values previously stored in Oldest are no
longer available. Table 4-5 illustrates calibration storage and the use of the CAL:DUMP? query
and CAL:COPY command.
TABLE 4-5. CALIBRATION STORAGE
COMMAND
WORKing
PRIor
OLDest
FACTory
MASTer
FIRst
1. CAL:SAVE
Cal 1 values
Factory cal.
Master cal.
First cal.
2. CAL:SAVE
Cal 2 values Cal 1 values
No Change
No Change
No Change
3. CAL:SAVE
Cal 3 values Cal 2 values Cal 1 values
No Change
No Change
No Change
4. CAL:SAVE
Cal 4 values Cal 3 values Cal 2 values
No Change
No Change
No Change
5. CAL:DUMP? FACTory;:CAL:COPY WORKing
Factory cal.
Cal 4 values Cal 3 values
No Change
No Change
No Change
6. CAL:DUMP? PRIor;:CAL:COPY WORKing
Cal 4 values
Factory cal.
Cal 4 values
No Change
No Change
No Change