BOP-1K-GL 022814

4-9

24.Set the BOP to maximum negative output current by sending

CAL:CURR MIN

. Continue to

measure the output current of the supply using the DVM connected to the sense resistor. To
adjust, send

CAL:DATA

commands as needed (see PAR. 4.3b) to adjust the BOP output

until the DVM reads as close as possible above (absolute value) the nominal full scale value
within tolerance specified in Table 4-4 for –FULL SCALE CURRENT.

25.Send

CAL:CPR MAX

to adjust the maximum positive current protection limit of the power

supply. To adjust, send

CAL:DATA

commands as needed (see PAR. 4.3c) to adjust the BOP

output until the DVM reads as close as possible above the nominal full scale value within tol-
erance specified in Table 4-4 for +FULL SCALE CPR LIMIT.

26.Send

CAL:CPR MIN

to adjust the minimum (maximum negative) current protection limit of

the power supply. To adjust, send

CAL:DATA

commands as needed (see PAR. 4.3c) to

adjust the BOP output until the DVM reads as close as possible above (absolute value) the
nominal full scale value within tolerance specified in Table 4-4 for –FULL SCALE CPR LIMIT.

27.Send

CAL:ZERO

to set the BOP output to zero. Connect 0V ±0.1mV d-c reference voltage to

external reference pin 11 referenced to pin 10 (SGND) of the BOP analog I/O port (A2A5J6).
Set the BOP to zero volts across the sense resistor (corresponding to zero current) by send-
ing

CAL:CEXT ZERO

. Send

CAL:DATA

commands as needed (see PAR. 4.3a) until the

reading is as close to zero as possible within the limits specified in Table 4-4 for CURRENT
ZERO.

28.Replace the 0V d-c reference applied to analog I/O port A2A5J6 pin 11 with a +10V ±0.1mV

d-c reference voltage. Set the BOP to maximum positive output current by sending

CAL:CEXT MAX

. Measure the current by reading the voltage across the sense resistor. To

adjust, send

CAL:DATA

commands as needed (see PAR. 4.3b) to adjust the BOP output

until the DVM reads as close as possible above the nominal full scale value within tolerance
specified in Table 4-4 for +FULL SCALE CURRENT.

29.Replace the +10V d-c reference applied to analog I/O port A2A5J6 pin 11 with a 0V ±0.1mV

d-c reference voltage. Set the BOP to zero volts across the sense resistor (corresponding to
zero current) by sending

CAL:CGA ZERO

. Send

CAL:DATA

commands as needed (see

PAR. 4.3a) until the reading is as close to zero as possible within the limits specified in Table
4-4 for CURRENT ZERO.

30.Replace the 0V d-c reference applied to analog I/O port A2A5J6 pin 11 with a +10V ±0.1mV

d-c reference voltage. Set the BOP to maximum positive output current by sending

CAL:CGA

MAX

. Measure the current by reading the voltage across the sense resistor. To adjust, send

CAL:DATA

commands as needed (see PAR. 4.3a) to adjust the BOP output until the DVM

reads as close as close to zero as possible within the limits specified in Table 4-4 for CUR-
RENT ZERO.

31.Send

CAL:ZERO

to prepare for calibration of IOUT_M_UNIT (master output) control signal.

Connect the DVM to pin 6 (IOUT_M_UNIT) referenced to pin 1 (SGND) of the PAR/SER
CONTROL IN connector (A2A5J3). Set the BOP to zero output current by sending

CAL:IOUT ZERO

. Send

CAL:DATA

commands as needed (see PAR. 4.3b) to adjust the

BOP output until the DVM reads 0V ±1mV.