B.31 [source:]current[:level]:protect:mode command, B.32 [source:]current[:level]:protect:mode? query, B.31 – KEPCO BOP-GL 1KW Operator Manual Firmware Ver.3.05 and higher User Manual
Page 139: Sour, B.32, B.33, B.34, B.35, B.36, B.37
![background image](/manuals/367930/139/background.png)
BOP-1KW-GL 022814
B-11
B.31 [SOURce:]CURRent[:LEVel]:PROTect:MODE COMMAND
CURR:PROT:MODE
Syntax:
Short Form: [SOUR:]CURR[:LEV]:PROT:MODE (EXT | FIX | LESS |)
Long Form: [SOURce:]CURRent[:LEVel]:PROTect:MODE (EXTernal | FIXed | LESSer |)
Description: Determines how current protection limits are controlled.
FIXED - Allows the limits to be controlled by digital signals from SCPI commands.
EXTERNAL: Allows the limits to be controlled by analog signals applied to the analog port; limits set
from SCPI command have no effect (see PAR. 3.3.6.1).
LESSER LIMIT: Allows protect limit to be automatically selected from either 1) the external analog limit
CURRent applied to the Analog I/O port or 2) the value set SCPI command). Whichever limit has a
lower absolute value (closest to zero) has effect.
B.32 [SOURce:]CURRent[:LEVel]:PROTect:MODE? QUERY
CURR:PROT:MODE?
Syntax:
Short Form: [SOUR:]CURR[:LEV]:PROT:MODE?
Long Form: [SOURce:]CURRent[:LEVel]:PROTect:MODE?
Returns FIXED, EXTERNAL or LESS
Description: Identifies how current protection limits are controlled. (See PAR. B.31 for details).
B.33 [SOURce:]CURRent[:LEVel]:PROTect:NEGative COMMAND
CURR:PROT:NEG
Syntax:
Short Form: [SOUR:]CURR[:LEV]:PROT:NEG
Long Form: [SOURce:]CURRent[:LEVel]:PROTect:NEGative
where
current.
Description: Establishes the negative protection limit for current, maximum is 1% greater than rated (nominal) out-
put current.
B.34 [SOURce:]CURRent[:LEVel]:PROTect:NEGative? QUERY
CURR:PROT:NEG?
Syntax:
Short Form: [SOUR:]CURR[:LEV]:PROT:NEG?
Long Form: [SOURce:]CURRent[:LEVel]:PROTect:NEGative?
Returns
Description: Identifies the negative protection limit for current.
B.35 [SOURce:]CURRent[:LEVel]:PROTect:POSitive COMMAND
CURR:PROT:POS
Syntax:
Short Form: [SOUR:]CURR[:LEV]:PROT:POS
Long Form: [SOURce:]CURRent[:LEVel]:PROTect:POSitive
where
current.
Description: Establishes the positive protection limit for current, maximum is 1% greater than rated output current.
B.36 [SOURce:]CURRent[:LEVel]:PROTect:POSitive? QUERY
CURR:PROT:POS?
Syntax:
Short Form: [SOUR:]CURR[:LEV]:PROT:POS?
Long Form: [SOURce:]CURRent[:LEVel]:PROTect:POSitive?
Returns
Description: Identifies the positive protection limit for current.
B.37 [SOURce:]CURRent[:LEVel]:PROTect:LIMit[:BOTH] COMMAND
CURR:PROT:LIM[:BOTH]
Syntax:
Short Form: [SOUR:]CURR[:LEV]:PROT:LIM[:BOTH]
Long Form: [SOURce:]CURRent[:LEVel]:PROTect:LIMit:[BOTH
where
current.
Description: Establishes the maximum value possible for both the positive and negative protection limits for cur-
rent. Requires MEM:UPD (PAR. B.12) to save for power up. See PAR. 3.5.5.2 and Figure 3-6 for spe-
cial programming considerations.