Current calibration measurements and tolerances -4 – KEPCO BOP-GL 1KW Operator Manual Firmware Ver.3.05 and higher User Manual
Page 114
![background image](/manuals/367930/114/background.png)
4-4
BOP-1K-GL 022814
a. ZERO Calibrations, main channel (voltage or current) (always done first). The only
means of adjustment is the
CAL:DATA
command which provides a total of 4095
increments of adjustment starting at 0, with 2047 increments in either direction to achieve maxi-
mum output levels of ±2.5% of E
Onom
or I
Onom
. Adjust the output to obtain the closest value to
zero.
b. MAX and MIN Calibrations, main channel (voltage or current). The primary means of
adjustment is the
CAL:DATA
command to either increase (
CAL:DATA 1
) or
decrease (
CAL:DATA -1
) the output value. Adjustment can be done in larger increments by
increasing the numeric value, e.g.,
CAL:DATA +10
or
CAL:DATA -10
. Adjust the output to
obtain the closest value above the full scale value.
c. VPR and CPR Calibrations, limit channels (positive and negative voltage or current
protection). The only means of adjustment is the
CAL:DATA
command which pro-
vides a total of 4095 increments of adjustment from zero to maximum. Adjust the output to
obtain the closest value above the full scale value.
TABLE 4-4. CURRENT CALIBRATION MEASUREMENTS AND TOLERANCES
MODEL
SENSE
RESISTOR
VALUE
(SEE NOTE AND
CURRENT
ZERO
±FULL SCALE
CURRENT
±FULL SCALE
CPR LIMIT
BOP 10-100GL
0.001 Ohm
0V
±0.01mV
0.100V (MAX)
–0.100V (MIN)
±0.01mV
0.100V (MAX)
–0.100V (MIN)
±0.05mV
BOP 20-50GL
0.001 Ohm
0V
±0.005mV
0.05V (MAX)
–0.05V (MIN)
±0.005mV
0.05V (MAX)
–0.05V (MIN)
±0.025mV
BOP 50-20GL
0.01 Ohm
0V
±0.02mV
0.2V (MAX)
–0.2V (MIN)
±0.02mV
0.2V (MAX)
–0.2V (MIN)
±0.10mV
REFER TO NOTES FOR SENSE
RESISTOR VALUE OTHER
THAN LISTED HERE
R
SENSE
SEE
NOTE.
SEE
NOTE A.
SEE NOTE B.
SEE NOTE C.
USE SPACE PROVIDED AT
RIGHT FOR CALCULATED
VALUES.
NOTES: The following formulas are used to calculate proper calibration values for Sense Resistors other than
those recommended.
R
SENSE
=
the measured value of the sense resistor in ohms (0.001% accuracy).
I
Onom
=
Nominal (rated) current output of BOP in Amperes (e.g., for BOP 50-20M,
I = 20A)
A. 0V, Tolerance = ±
(
R
SENSE) x (0.1) x (
I
Onom)--
result in mV.
B. (
R
SENSE
)
x (
I
Onom)--
result in V, Tolerance =
±(
R
SENSE) x (0.1) x (
I
Onom)--
result in mV.
C. (
R
SENSE
)
x (
I
Onom)--
result in V, Tolerance =
±(
R
SENSE) x (0.5) x (
I
Onom)--
result in mV.