GDS-800 Series Digital Storage Oscilloscope Operation Manual

48

z

Previous Menu: back to previous menu.

For Template Auto

z

Template Auto: The two “Go-No Go” templates are generated

automatically from the subject signal.

(1): According to the subject signal, the two templates (RefA and RefB) are

created by “Template Auto” automatically. The adjustable range (Tolerance)

is available from ±0.4% to ±40% (or ±0.04 division to ±4 division).

Figure 5-23: Template created “Template Auto” function

z

Source CH1/CH2: Select the input source of channel 1 or channel 2 as

“Go-No Go” template.

z

Tolerance %: Select the tolerance percentage (or division) range for

vertical and horizontal scales from the subject signal. The adjustable range

is available from ±0.4% to ±40% (or ±0.04 division to ±4 division).

z

Save & Create: Press F4 softkey to save the settings. In the meantime,

the original saved reference A or B from “Save/Recall” function will be also

changed as current setting

Previous Menu: Back to previous menu.