GW Instek GDS-800 Series User Manual User Manual
Page 47
GDS-800 Series Digital Storage Oscilloscope Operation Manual
47
Template Min: The minimum template is always selected from the
Reference B of “Save/Recall” function or from the one hundred sets of
custom-made waveforms.
Note: The maximum and minimum templates can not select the same
custom-made waveform source. For example, if the custom-made
waveform of “Template Max” is selected as number 10, the “Template Min”
have to choice the different custom-made waveform number except
number 10.
z
Source RefA/RefB/1~100: Indication for the source of maximum
template or minimum template (Reference A for maximum template;
Reference B for minimum template).
This oscilloscope can accept one hundred custom-made waveforms from a
personal computer with the advance “FreeCapture” software (this freeware
can be downloaded from our website:
http://www.goodwill.com.tw
). The
“FreeCapture” software can edit any specified waveforms from users and
input to oscilloscope as maximum (or minimum) templates. The maximum
custom-made waveforms input capacities are one hundred sets. For the
details of “FreeCapture” operations, please refer to the “FreeCapture”
operation manual.
User can select the different custom-made waveforms sets by rotating the
VARIABLE knob. The maximum or minimum template CAN NOT select the
identical custom-made waveforms location. The two maximum and
minimum must be selected form the different custom-made waveforms
location. Otherwise, the “Go- No Go” function will go wrong.
z
Position: Adjust the vertical position for maximum template or
minimum template or custom-made waveforms. The adjustable ranges are
from ±0.04 division to ±4 division.
z
Save: Press F4 softkey to save the current settings. In the meantime,
the original saved reference A or B from “Save/Recall” function will be also
changed as current setting.