beautypg.com

GW Instek AFG-100/200 Series User Manual User Manual

Page 54

background image

AFG-125/225 Series Options Manual

54

6. Select the N Cycle parameters from the side

menu:

Sets the number of burst cycles* :
1~ 65,535, infinite.

Sets the starting phase of the burst
waveform.

Sets the period* length in seconds
for the burst waveform.

Note

*The Period setting will affect the number of cycles
that can be chosen for any given frequency. The
relationship is shown in the following equation:

Burst Cycle < (Burst Period x Wave Frequency)

Trigger Settings

7. Press Trigger set from the side

menu.

8. Select the Trigger parameters from the side

menu:

Sets the trigger as internal. An
internal trigger is generated at the
end of each period.

Sets the trigger to Manual mode.
This Manual Trigger key must be
pressed each time to output a burst
waveform.

Sets a delay time between the
trigger and the burst waveform
output: 0 ~ 655350nS.

9. See page 29 to set the output

settings and to turn the output on.

Page 29