3.1 Conduit Connections

The Shark has four 1/2” conduit holes, 2 on each side of the enclosure as shown on fig. 2.1.
The unit is shipped with these holes plugged with liquid tight conduit seals. These must be left
in unused holes to maintain the NEMA 4X integrity. Use approved conduit hubs to connect the
conduit, connect these to the conduit before connecting to the enclosure.

Wire Specification: Size and fuse wire accroding to local electrical code. Maximum current not
to exceed 10A when used to power auxillary decvices powered via internal connections.

3.2 A.C. Power Connections

The Shark is available in two power models.
The Shark-240 is designed to operate at 240 VAC.
The Shark-120 is designed to operate at 120 VAC.
To connect power to the Shark, remove the terminal block plug P3 and connect the wiring as
shown below.

Section 3 - Electrical Connections and Setup

Page 6

S

H

A

R

K

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

Section 3 - Electrical Connections and Setup

Caution: This instrument uses 120 or 240 50/60 Hz AC power. Opening the enclosure
door exposes you to potentially hazardous line power voltage which might be pres-
ent on the terminals of plug P3 and P4. Always remove line power before working in
this area. If the relay contacts on P4 are powered from a seperate source from the
line power on P3, be sure to disconnect that power before proceeding. The Shark flip
out door contains only low voltage and is safe to handle.

AC POWER CONNECTIONS FOR 120 VAC

AC POWER CONNECTIONS FOR 240 VAC

LINE

LINE

GROUND

SHARK-240

NEUTRAL

LINE

GROUND

SHARK-120

L

L

P3

G

N

L

G

P3

Figure 3.1 A.C. Power Connections

Dwg# N104-33