beautypg.com

6 paddle wheel flow sensor connections and setup – Aquametrix SHARK-120/240 Controllers User Manual User Manual

Page 12

background image

Section 3 - Electrical Connections and Setup

Page 10

S

H

A

R

K

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

Section 3 - Electrical Connections and Setup

3.6 Paddle Wheel Flow Sensor connections and setup

P6

8

1 2 3 4

6

5

7

10

9

11 12 13 14 15 17

16

+12V (6)

GND (15)

(FLOW SWITCH INPUT) (13)

SIGNAL + (14)

SIGNAL - (GND) (15)

COLLECTOR

CONTACT

CLOSING THE FLOW SWITCH INPUT

TO GROUND WILL ZERO THE FLOW DISPLAY

DRY

OPEN

OR

9ma

+12V O.C.

FLOW SENSOR

EXTERNAL POWERED

FLOW SENSOR

SELF POWERED

16 17

15

14

13

12

11

9 10

8

7

5 6

4

3

2

1

SIGNAL - (15)

SIGNAL + (14)

SHIELD

SHIELD

SHIELD (15)

Figure 3.5 Connections for Flow Sensor

SETUP OF UNITS

OF VOLUME

SETUP OF UNITS

CALIBRATION FACTOR

ENTER FLOW SENSOR

SELECT FLOW METER

SEC. 7.6

OF TIME

SEC 7.5

SEC. 7.4

SEC. 7.1

TOTAL 0

TOTALIZER RESET

UNITS OF TIME

UNITS OF VOLUME

K FACTOR

METER SELECTION

RESET TOTALIZER

RUN MODE

TO ZERO

SEC 7.0

Note: Leave 4” to 6” slack for all wires connected to the ter-
minals of P6. Slack required so that wires do not interfere
with opening or closing of the front door.

Once connected, step through the LCD menus to select the
sensor in the order shown. The Sensor K factor (pulses per
U.S. Gallon) is usually printed on the side of the sensor or
on a label attached to the sensor cable.

The drawing shows the connnections for a typical paddle wheel flow sensor. The cable to the sensor
should not exceed 2000’ (600 meters).
The Shark controller also supports the use of an external “flow switch”. When the flow switch input is
grounded, either through a dry contact or solid state input, the flow display will be held at zero. This is
useful to ensure the flow reading remains locked at zero when conditions require it. The flow will start
reading again when the input is opened. If the flow switch function is not desired, simply leave it discon-
nected and the flow meter will read as normal.

Dwg# N104-37

CAUTION:
Always remove line power before unplugging or
plugging in the P6 connector