Aquametrix SHARK-120/240 Controllers User Manual User Manual
Page 11
S
H
A
R
K
MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL
Section 3 - Electrical Connections and Setup
Section 3 - Electrical Connections and Setup
Page 9
3.5 Conductivity Cell (Contacting style) connections and setup
WHITE (CELL) (1)
BLACK (CELL) (4)
RED (TEMP. SENSOR) (4)
GREEN (TEMP. SENSOR) (5)
P6
8
1 2 3 4
6
5
7
10
9
11 12 13 14 15 17
16
CONTACTING
CONDUCTIVITY
CELL
Figure 3.4 Connections for Conductivity Cells
SET TEMPERATURE COMPENSATION
3Kohm NTC (thermistor)
1Kohm RTD
ENSURE SENSOR IS CORRECT TYPE
SEC. 6.7
SELECT CONDUCTIVITY RANGE
SELECT COND METER
SEC. 6.13
OR
FOR PROCESS
SEC 6.9
SEC. 6.6
1000uS 25.0C
MANUAL CAL COND
TEMP COMP CURVE
TEMP SENSOR
COND RANGE
METER SELECTION
CALIBRATE WITH
RUN MODE
REFERENCE SOLUTIONS
SEC 6.1
DRY CAL COND
CALIBRATE WITH FACTORY
SPECIFIED CELL CONSTANT
SEC 6.2
Once connnected, step
through the LCD menus to
select the cell in the order
shown. The TEMP COMP
CURVE setup default is
1.8%/deg C. This is accept-
able for most process appli-
cations. If your process is
significantly different from
this, change the setting in
the TEMP COMP CURVE
menu.
The drawing shows the connections for the Aquametrix Conductivity Cells (Contacting style). The cable
should be run in a conduit seperate from the AC power wires, and via a seperate conduit hole. The cell
cable length should not exceed 300ft. (91 meters).
Note: Leave 4” to 6”
slack for all wires
connected to the
terminals of P6.
Slack required so
that wires do not
interfere with open-
ing or closing of the
front door.
Dwg# N104-36
CAUTION:
Always remove line
power before
unplugging or
plugging in the P6
connector