beautypg.com

Aquametrix SHARK-120/240 Controllers User Manual User Manual

Page 11

background image

S

H

A

R

K

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

Section 3 - Electrical Connections and Setup

Section 3 - Electrical Connections and Setup

Page 9

3.5 Conductivity Cell (Contacting style) connections and setup

WHITE (CELL) (1)

BLACK (CELL) (4)

RED (TEMP. SENSOR) (4)

GREEN (TEMP. SENSOR) (5)

P6

8

1 2 3 4

6

5

7

10

9

11 12 13 14 15 17

16

CONTACTING

CONDUCTIVITY

CELL

Figure 3.4 Connections for Conductivity Cells

SET TEMPERATURE COMPENSATION

3Kohm NTC (thermistor)
1Kohm RTD

ENSURE SENSOR IS CORRECT TYPE

SEC. 6.7

SELECT CONDUCTIVITY RANGE

SELECT COND METER

SEC. 6.13

OR

FOR PROCESS

SEC 6.9

SEC. 6.6

1000uS 25.0C

MANUAL CAL COND

TEMP COMP CURVE

TEMP SENSOR

COND RANGE

METER SELECTION

CALIBRATE WITH

RUN MODE

REFERENCE SOLUTIONS

SEC 6.1

DRY CAL COND

CALIBRATE WITH FACTORY

SPECIFIED CELL CONSTANT

SEC 6.2

Once connnected, step
through the LCD menus to
select the cell in the order
shown. The TEMP COMP
CURVE setup default is
1.8%/deg C. This is accept-
able for most process appli-
cations. If your process is
significantly different from
this, change the setting in
the TEMP COMP CURVE
menu.

The drawing shows the connections for the Aquametrix Conductivity Cells (Contacting style). The cable
should be run in a conduit seperate from the AC power wires, and via a seperate conduit hole. The cell
cable length should not exceed 300ft. (91 meters).

Note: Leave 4” to 6”
slack for all wires
connected to the
terminals of P6.
Slack required so
that wires do not
interfere with open-
ing or closing of the
front door.

Dwg# N104-36

CAUTION:
Always remove line
power before
unplugging or
plugging in the P6
connector