<6. CALORIE / PARAMETER SETTING>

6-9

IM 11T3E1-01E

(2) Density/Calorie Conversion Factor Setting Mode
In order to this mode, select “MAN.CAL” in before page, and press [YES] key.

This mode is used to set a factor for converting density to calorie. This setting is mandatory when

measuring the calorie.

(a) Automatic zero-point adjustment [AT_ADJ] : 0 to 133.000, 0 to 3.5000
(b) Factor setting [SET_K] :
• Zero point [K0_ADJ] : -99999 to 99999
• K0 [K0] :

-99999 to 99999

• K1 [K1] :

0 to 99999

Note: The result of executing item (a) above is also reflected on [K0_ADJ] in item (b).

[ENT]

[ENT]

ADJUST

XXX.XXX

VALVE

END

X.XXXX

K0_ADJ

XXX.XXX

CAL.SET

X.XXXX

K0

X.XXXX

K1

Measurement mode

[YES]

[NO]

[NO]

[YES]

[ENT]

AT_ADJ

SET_K

[YES]

[YES]

[ENT]

[ENT]

F0602.ai

*2: Using the [ ], [ ] and [ENT] keys, set a value (numeral position of the decimal point and sign).

Each press of the [ ] key moves through the fields of the numeral decimal point and sign in this order.

If you enter a value that exceeds the limits, the display shows [*OVER].

Press the [YES] or [NO] key to re-set the value.

*3: Set a value using the [ ], [ ] and [ENT] keys.

*4 When the reading stabilizes, press the [ENT] key to confirm the factor.

This parameter does not accept input from the [ ] and [ ] keys.

[NO]

MAN.CAL

[YES]

ZERO

[NO]

SPAN

[NO]

*1: See "Figure 6.1 Modes of Operation Level" on page 6-8.

*

3

*

4

*

2

*

2

*

2

*

1

Figure 6.2 Density / Calorie Conversion Factor Setting Mode