<5. DENSITY / PARAMETER SETTING>

5-14

IM 11T3E1-01E

2. Alarm-point Setting Mode [*ALARM]
Set the upper /lower limits for the measured values of the parameters noted below, for the pur-

pose of raising an alarm.

(a) Select a parameter for which alarms are raised.
・ Physical density [*DENS]
・ Compensated density [*C_DNS]
・ Specific gravity [*SP_GR]
・ Calorific value [*CALRY]
・ Molecular weight [*MOL]
・ Concentration [*CONCT]
Press the [YES] key to enter the desired mode when it is indicated.

(b) Set the upper/lower limits.
・ Physical density [*L_DNS], [*H_DNS]

Configure range:

00.0000 to 60.0000 (kg/m

3

),

0.00000 to 4.00000 (lb/ft

3

)

・ Compensated density [*L_CP.D], [*H_CP.D]

Configure range:

0.0000 to 6.0000 (kg/m

3

),

0.00000 to 0.40000 (lb/ft

3

)

・ Specific gravity [*L_SPC], [*H_SPC]

Configure range:

0.0000 to 5.0000

・ Calorific Value [*L_CAL], [*H_CAL]

Configure range:

000.000 to 133.000 (MJ/m

3

),

0.0000 to 3.5000 (kBTU/ft

3

)

・ Molecular weight [*L_MOL], [*H_MOL]

Configure range:

000.00 to 140.00

・ Concentration [*L_CON], [*H_CON]

Configure range:

000.0 to 100.0