Yokogawa GD40 Gas Density Detector User Manual
Page 113
![background image](/manuals/360159/113/background.png)
<6. CALORIE / PARAMETER SETTING>
6-24
IM 11T3E1-01E
CODE 15: Setting of Data for Automatic/Semi-automatic Calibration
This setting defines the data needed to carry out automatic/semi-automatic calibration.
(a) Selection of calibration item [*CAL.P]
・ Zero and span:
0
・ Zero:
1
・ Span:
2
(b) Setting of calibration time [*CAL.T]
Configurable range: 00 to 60 (unit : minute)
(c) Setting of stabilization time [*STAB.T]
Configurable range: 00 to 60 (unit : minute)
(d) Setting of calibration starting time
(1) Setting of year/month/day [*Y_M_D]
Configurable range: 00.01.01 to 99.12.31
(2) Setting of hour/minute [*H_M]
Configurable range: 00.00 to 23.59
(e) Selection of unit of calibration interval [*CYCL.U]
・ Hour: 0
・ Day: 1
(f) Selection of calibration interval [*CYCL.T]
Item (e) above is set to "0".
Configurable range: 00 to 23 (for hour)
Item (e) above is set to "1".
Configurable range: 000 to 255 (for day)
For more information, see Chapter 9, “Calibration Procedure”.