beautypg.com

3 automatic calibration, Automatic calibration -7 – Yokogawa GD40 Gas Density Detector User Manual

Page 166

background image

<9. CALORIE / CALIBRATION PROCEDURE>

9-7

IM 11T3E1-01E

9.3.3

Automatic Calibration

This mode of calibration takes effect when you select “enable” for automatic calibration in CODE

13 - [*AUTO.C] - at the service level. Even during automatic calibration, you can interrupt the

system with the semi-automatic or manual mode of calibration. One cycle of automatic calibration

is skipped, however, if the time of automatic calibration arrives when semi-automatic calibration is

in progress.

*1 : Pressing the [MODE] key, change it to display [WAIT] indication.
*2 : Pressing the [MODE] key, once again returns it measuring mode.

X.XXXX

ZERO

X.XXXX

SPAN

X.XXXX

WAIT

* 1

* 1

* 2

F0906.ai

Measurement mode

Measurement mode

Figure 9.5 Automatic Calibration Mode

This manual is related to the following products: