beautypg.com

Yokogawa GD40 Gas Density Detector User Manual

Page 76

background image

<5. DENSITY / PARAMETER SETTING>

5-21

IM 11T3E1-01E

CODE 11: Setting for Compensated Density Measurement

This setting defines the reference temperature and pressure required to obtain the compensated

density.

(a) Setting of reference temperature

[

*C.D.TMP

]

Configurable range:

-20.0 to 80.0 (for °C)

-4.0 to 176.0 (for °F )

(b) Setting of reference pressure

[

*C.D.PRS

]

Configurable range:

0.10 to 999.99 (for kPa),

0.0001 to 9.9999 (for MPa),

0.015 to 99.999 (for psi )

11

*CODE

*SERVC

XX.X

*C.D.TMP

XXX.XX

*C.D.PRS

[YES]

F0513.ai

*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.

*2: Enter the code number.

*3: Set a value using the [ ], [ ] and [ENT] keys.

If you enter a value that exceeds the limits, the display shows [*OVER].

Press the [YES] or [NO] key to re-set the value.

*4: The position of the decimal point changes depending on the setting of CODE 23.

If the code is set to "0" (°C), the reading is XX.X;

if the code is set to "1" (°F), the reading is XXX.X.

[ENT]

[ENT]

[ENT]

*5: The position of the decimal point changes depending on the setting of CODE 20.

If the code is set to "0" (kPa), the reading is XXX.XX;

if the code is set to "1" (MPa), the reading is X.XXXX;

if the code is set to "2" (psi), the reading is XX.XXX.

*

3

*

5

*

3

*

4

*

2

*

1

Figure 5.12 Setting for Compensated Density Measurement

This manual is related to the following products: