<7. HYDROGEN PURITY / PARAMETER SETTING>
7-25
IM 11T3E1-01E
CODE 31: Showing/Hiding of Negative Measured Values [*MINUS]
This setting determines whether negative measured values (-) should be shown or hidden.
・ Show: 0
・ Hide: 1
[ENT]
*CODE
31
*SERVC
X
*MINUS
[ENT]
[YES]
F0720.ai
*
1
*
2
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
*
3
Figure 7.20 Showing/Hiding of Negative Measured Values
CODE 40: Indication of Calibration Coefficients (Read-only)
This setting determines the calibration coefficients to be indicated.
(a) Zero [*C_K_Z]:
indicated as SX.XXXX
(b) Span [*C_K_S]:
indicated as X.XXXX
[ENT]
40
*CODE
*SERVC
SX.XXXX
*C_K_Z
X.XXXX
*C_K_S
[ENT]
[ENT]
[YES]
F0721.ai
*
1
*
2
*
3
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: By pressing the [ENT] key, the next indication is displayed.
This parameter does not accept the [ ] and [ ] key input.
Figure 7.21 Indication of Calibration Coefficients